196 / 196 M electronic IP65
14
en
Calibration in 4 positions:
Calibration in 4 positions is needed if the following is
displayed:
1. The reverse test angle is ≥ 0.1° to the normal position
--> too great a deviation.
2. Change in internal reference
3. Change in temperature since last calibration.
The electronic spirit level is calibrated in 4 measuring positions
consecutively, turning 90° / 180° each time.
Calibration in 4 positions: Step 1
Press CAL and MODE buttons at the same time. Align the
electronic spirit level accurately against a wall and press
the CAL button to confirm.
Only the 1st calibration step is performed with the
vial.
1x
Flashing segments indicate the positions still to be calibrated.
Non-flashing segments indicate the positions already
calibrated.