St. Jude Medical Angio-Seal Evolution Скачать руководство пользователя страница 5

4

Conclusions:

Results of the clinical study demonstrate that patients that have undergone diagnostic angiography and have received a 6F Angio-Seal Device can safely and effectively ambulate in less 
than 20 minutes and be discharged one hour post ambulation.

HOW SUPPLIED

The Angio-Seal Evolution Vascular Closure Device is supplied sterile in a bag. This bag contains the following supplies.
1 each:
• Insertion 

Sheath

• Arteriotomy 

Locator 

•  6F (2.0 mm) – 0.035 in. (0.89 mm) Guidewire with J-Straightener
 or
 

8F (2.7 mm) – 0.038 in. (0.96 mm) Guidewire with J-Straightener

•  Angio-Seal Device (refer to device component figure)

 Device 

Components 

 

 

Device 

Sleeve

Reference

Indicator

Suture Release

Button

Device Handle

Anchor

Suture

Device Handle

Device 

Sleeve

Compaction 

Tube

Hemostatic 

Collagen 

Sponge

Bypass 

Tube

Suture Release 

Button

Reference 

Indicator

Carrier 

Tube

ANGIO-SEAL DEVICE INSERTION PROCEDURE

The medical techniques and procedures described in these Instructions for Use do not represent ALL medically acceptable protocols, nor are they intended as a substitute for the 
clinician’s experience and judgment in treating any speci

fi

 c patient.

The Angio-Seal procedure is composed of three stages:

A. Locate the Artery
B. Set the Anchor
C.  Seal the Puncture

A. Locate the Artery

1.  Assess the puncture site location and evaluate the femoral artery characteristics prior to placing the Angio-Seal Device by injecting contrast medium through the procedure sheath 

followed by an angiogram.

2.  Using sterile technique, remove the Angio-Seal Device contents from the foil package, taking care to pull foil apart completely before removing the Angio-Seal Device.

NOTE: The Angio-Seal Device must be used within one hour after opening the foil pouch due to the moisture-sensitive nature of the product.

3.  Insert the arteriotomy locator into the Angio-Seal insertion sheath (Figure 1), making sure the two pieces snap together securely. To ensure proper orientation of the arteriotomy 

locator with the sheath, the hub of the locator and the sheath cap fit together only in the correct position. The reference indicator on the locator hub must align with the reference 
indicator on the sheath cap.

Figure 1

Reference 

Indicator

Insertion 

Sheath 

Cap

Drip Hole 

Blood Inlet 

Hole

Insertion 

Sheath

Arteriotomy 

Locator

1.5 cm

4.  Insert the Angio-Seal guidewire into the procedure sheath that is currently in the patient. If the procedure sheath is smaller than the Angio-Seal sheath, it is advisable to ensure that 

the skin incision is of sufficient size to accommodate the Angio-Seal insertion sheath.

5.  Remove the procedure sheath, leaving the guidewire in place to maintain vascular access.
6.  Thread the Angio-Seal arteriotomy locator/insertion sheath assembly over the guidewire; the drip hole (located above the arteriotomy locator hub) will be oriented down and away so 

the flow of blood can be observed. Make sure the reference indicator on the insertion sheath is facing up, insert the assembly into the puncture tract. When the tip of the insertion 
sheath is about 1.5 cm into the artery, blood will begin to flow from the drip hole in the locator (Figure 2).

Содержание Angio-Seal Evolution

Страница 1: ...INSTRUCTIONS FOR USE Angio Seal Evolution Vascular Closure Device ...

Страница 2: ...f the puncture site is proximal to the inguinal ligament as this may result in a retroperitoneal hematoma PRECAUTIONS Special Patient Populations The safety and effectiveness of the Angio Seal Device has not been established in the following patient populations Patients undergoing an interventional procedure who are being treated with warfarin Patients who have known allergies to beef products col...

Страница 3: ...infection at the puncture site should be taken seriously and the patient monitored carefully Surgical removal of the device should be considered whenever an access site infection is suspected Collagen deposition into the artery or thrombosis at puncture site If this condition is suspected the diagnosis can be confirmed by duplex ultrasound Treatment of this event may include thrombolysis percutane...

Страница 4: ...gnostic angioplasty procedures using a 6 French or smaller procedural sheath The Angio Seal STS was evaluated for safety and efficacy when ambulating patients as soon as possible following device deployment and discharging patients as early as 30 minutes following ambulation Methods Patients who met all inclusion and no exclusion criteria were enrolled during a pre procedure screening evaluation F...

Страница 5: ...an angiogram 2 Using sterile technique remove the Angio Seal Device contents from the foil package taking care to pull foil apart completely before removing the Angio Seal Device NOTE The Angio Seal Device must be used within one hour after opening the foil pouch due to the moisture sensitive nature of the product 3 Insert the arteriotomy locator into the Angio Seal insertion sheath Figure 1 makin...

Страница 6: ...gure 3 Figure 3 8 From this point advance the arteriotomy locator insertion sheath assembly until blood begins to flow from the drip hole on the locator NOTE Over insertion of the arteriotomy locator insertion sheath assembly into the artery beyond 2 cm may increase the chance of premature anchor hook up or interfere with the anchor s performance to achieve hemostasis If blood flow does not resume...

Страница 7: ...on Figure 7 Reference Indicator Carrier Tube Ensure tab is in rear holding position prior to insertion Bypass Tube Hemostatic Valve 2 Confirm that the reference indicator on the insertion sheath is facing up To ensure proper orientation of the Angio Seal Device with the sheath the sheath cap and the device sleeve only fit together in the correct position The reference indicator on the device handl...

Страница 8: ...ble NOTE If the device sleeve separates from the sheath while attempting full rear lock positioning do not push the Angio Seal device forward to reattach to the sheath cap Complete sealing of the puncture following instructions in C 1 Figure 10 6 Incorrect Indicator Alignment Distal end of device handle completely covers the white band on the device sleeve Figure 11 If anchor catches prematurely a...

Страница 9: ...Pull back on the device handle at the angle of the puncture tract maintaining a steady uninterrupted motion Figure 13 until the colored compaction marker is revealed Figure 14 Pause and assess for hemostasis NOTE If hemostasis is not achieved check to see if the colored compaction marker is exposed If not repeat this step Figure 13 Figure 14 P u l l Compaction Tube Collagen Tissue Artery Sheath Ma...

Страница 10: ...ct fails to perform within normal tolerances for a patient due to a defect in materials or workmanship SJM will provide at no charge a replacement SJM product for the patient s use This limited warranty applies only if each of the following conditions is met 1 The product was packaged and labeled by SJM 2 The failed product must be returned to SJM and becomes the property of SJM 3 The product has ...

Страница 11: ...RILE R Sterilized by gamma radiation Do not use if temperature indicator dot on package has turned from light grey to dark grey or black Prescription Use Only Manufacture Date Manufacturer Quantity REF Reorder Number Lot Number Vascular Closure Device Vascular Closure Device Accessory Guidewire Product of Ireland Accessory Guidewire Product of Ireland SYMBOLS ...

Страница 12: ...NTROL LESS RISK are trademarks and service marks of St Jude Medical Inc and its related companies 2011 St Jude Medical All rights reserved St Jude Medical 14901 DeVeau Place Minnetonka MN 55345 2126 USA Phone 1 888 864 7444 1 952 933 4700 sjm com ART100041600 Ver A 04 2011 ...

Отзывы: