Single Gas Analyzer (SGA)
Operations Manual
Super Systems Inc.
Page 25 of 44
Page 25 of 51
Instrument Configuration
The Instrument Configuration page allows you to set values for the following information:
Set Date/Time: This option, when pressed, will sync the current time of the main board
to the computer on which the web interface is running.
Web Access Code: This allows you to set the required passcode to access the web
interface.
Min. Gas %: This reading indicates the point at which anything below will be read as zero
for internal calculations.
N2 Flow: The value entered here will be used in nitriding calculations. Enter the
applicable N2 flow in your system.
NH3 Flow: The value entered here will be used in nitriding calculations. Enter the
applicable NH3 flow in your system.
DA Flow: The value entered here will be used in nitriding calculations. Enter the
applicable DA flow in your system.
H2 Flow: The value entered here will be used in nitriding calculations. Enter the
applicable H2 flow in your system.
CO2 Pres: (0-10) – Any value entered here will be used in place of the actual CO2 reading
for internal calculations. (Zero is the default setting)
Suppress Negative – When activated, any negative readings will be treated as zero.