Instruction Manual 

 

Static Field Meter STATIRON 

DZ4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Thank  you  for  purchasing  the  Statiron  DZ4.  Please  read  this  manual  before  using  the 
product  in  order  to  fully  understand  its  functions.  Also  make  sure  to  store  this  manual  so 
that it can be referred to in the future. 

 

Functions 

The static field meter Statiron DZ4 is a portable static electrical potential meter with digital readout, for measuring 
the electrical potential of charged objects. 
When  measuring  fluctuating  static  electricity,  you  can  temporarily  hold  an  indicated  value,  and  check  maximum 
voltages in the MAX mode. 
 

Characteristics 

1. Enables you to measure surface potentials of charged bodies without coming into contact. 
2. Enables you to take measurements in narrow locations by rotating the sensor head. 
3. With the optional ion balance plate, you can carry out ionizer balance checks. 

 

 

SHISHIDO ELECTROSTATIC, LTD.

Содержание Statiron DZ4

Страница 1: ...lectrical potential meter with digital readout for measuring the electrical potential of charged objects When measuring fluctuating static electricity you can temporarily hold an indicated value and c...

Страница 2: ...osed to water or oil Maintenance Regularly remove any built up dirt etc from the optional ion balance plate Built up dirt can cause insulation faults Make sure to turn the main power of the device OFF...

Страница 3: ...using it for the first time Instruction manual warranty x 1 this document Main Device Earth Lead x 1 Strap x 1 Battery 006P type 9V alkaline dry cell battery x 1 Soft case Model number ODZ4 SFTCS x 1...

Страница 4: ...al changes according to the field strength concentrated in the electrode of the sensor portion of the meter and the size of the charged object In particular if the charged object is smaller than the t...

Страница 5: ...ad to a grounded object To obtain accurate measurements the earth terminal must be grounded Measurement Procedure Measuring Static Electricity 1 Turn the meter on Press the POWER button for more than...

Страница 6: ...re 3 Fig 3 Red LED mark NOTE If the recommended measuring distance of 30mm cannot be attained do not bring the meter any closer if the potential displayed exceeds 19 99kV Doing so may cause damage to...

Страница 7: ...e the I B mode 3 Zero point adjust the meter Point the sensor at a grounded object and press the 0ADJ button NOTE Zero point adjustment is reset when you turn the meter off 4 Attach the ion balance pl...

Страница 8: ...s occurs stop measuring immediately as this may cause damage to the meter 7 The value on the display panel is the measurement result unit kV 8 In I B mode you can temporarily hold the measured value b...

Страница 9: ...ed on the upper right of the display panel When the battery charge icon shows one unit left replace the battery Error Display The signal detection method of the meter is an oscillating chopper method...

Страница 10: ...correct Remedy 1 Perform zero point adjustment again Cause 2 Structural components close to the sensor are charged Remedy 2 Wait until the charge in the components has attenuated The display does not...

Страница 11: ...measured and sensor Measuring distance adjustment Red LED beam focusing method Sensor head rotation angle 180 stops every 45 Display LCD display with built in backlight Polarity display polarity pola...

Страница 12: ...where you purchased the product SHISHIDO ELECTROSTATIC LTD http www shishido esd co jp HEAD OFFICE MARUNOUCHI BLDG 9F 918 4 1 MARUNOUCHI 2 CHOME CHIYODA KU TOKYO 100 6309 SALES DEPARTMENT TOKYO BRANCH...