CHARACTERISTICS 

SPECIFICATIONS 

Size 12

1/2"L x 4"H x 4"W  (32cm L x 11cm H x 11cm W) 

Weight 

2.6 lbs  (1.18 kg) 

Vacuum Pump 

12V DC, 0.7 A.  Exceeds 500mmHg.   
Lower levels of nega ve pressure will be observed at al tude. 

Regulator  

Reduces vacuum from 

500+mmHg (High Vacuum / Low Flow) to approximately  

85mmHg (Low Vacuum / Low Flow).  Typically 

70mmHg to 

100mmHg but the range 

can be larger due to condi ons such as variances in ba ery charge and condi on. 

Ba ery  

12V DC 
Sealed Lead Acid replacement part #80611

100 

Alkaline ba ery replacement part #80613

100 

Please read page 6 for ba ery care sugges ons 

Collec on Canister 
(13) 

300cc/ml capacity. Non Sterile, Single Use  
Model 2480: Standard replacement Quickdraw canister (shown on page 6) 
Model 2488: Barbed Quickdraw canister (shown on front cover)

 

Pa ent tubing 

Vinyl tubing 9/32”ID x 36”L (7.1mm ID x 91.44 cm L) 
SSCOR part #43203 

Suc on Tip 

HI

D® “Big S ck®” SSCOR part #44241C 

SSCOR S3 SSCOR Part #44305C 

PROBLEM 

POSSIBLE CAUSE 

CORRECTIVE ACTION 

Does not func on 
when switch is 
turned on (DC Power) 

Ba ery  discharged 
 
Molex connec ons disconnected 
Damaged PC Board 

Reconnect to charging source to ac vate charger 
or install a new fully charged ba ery 
Open unit and re

connect Molex connec ons 

Replace PC Board 

Ba ery Condi on  
indicator lights not lit 

Power cord disconnected 
Damaged PC Board 

Reconnect cord 
Replace PC Board 

No suc on when 
pump is running 

Canister not secure to chassis 
Regulator vent is open 
Tip cover is on the catheter 
Fluids have shut down the 

fi

lter in 

the canister 

Latch canister to chassis 
Replace cap on vent 
Remove  p cover from catheter 
Replace canister 

Quickdraw - Model 2400 Series 

Clearing The Airway Is Our #1 Priority 

General Specifications 

Trouble Shooting 

Rev. M (02/17) 

Содержание Quickdraw 2400 Series

Страница 1: ... Airway Is Our 1 Priority Operating Instructions Maintenance Manual Tel 1 818 504 4054 Tel 1 800 434 5211 INC 11064 Randall Street Sun Valley CA 91352 USA www sscor com shop sscor com Email marketing sscor com techsupport sscor com ...

Страница 2: ...ut or other connectors has formed a system and is there fore responsible for the system to comply with the requirements of IEC 60601 1 1 If in doubt contact a qualified technician or your local representa ve For further informa on reference the SSCOR Technical Manual 6 This device is not intended for suc oning neonates 7 To prevent fire of injury when ba eries are not in the device or charger alwa...

Страница 3: ... 140 F or 60 C may cause deforma on of the ba ery case and damage the ba ery Sealed lead acid ba eries can easily be maintained to permit proper opera on of the equipment To protect the ba ery a er each procedure turn the unit off put the unit on charge and always store the suc on device at room temperature The only way to assure the ba ery has func onal capacity even if it is indicated the ba ery ...

Страница 4: ...anister Single Use Only Serialized Label A en on Consult Accompanying Documents Serialized Label Temperature Limita ons Serialized Label Separate collec on for electronic equipment Serialized Label Type BF Equipment Shipping Carton Manufacturer Shipping Carton Date of Manufacture Shipping Carton Authorized Representa ve in the European Community Quickdraw Model 2400 Series 4 Clearing The Airway Is...

Страница 5: ...the unit on charge and always store the pump at room temperature Two Posi on Regulator If full power 500 mmHg is not required the units low se ng is available by removing the regulator vent cap from the regulator vent 4 Be sure to replace the cap on the vent when full power is required or to dispose of the canister DC Ba ery Charging from Vehicle To charge the Quickdraw from the vehicle s DC elect...

Страница 6: ...ility Reverse the posi on of the canister and slide the body of the canister into the rails and the catheter p through the reten on ring Push the canister toward the chassis un l the canister latch on top of the unit snaps into the detent on the canister To release the canister pull the canister latch up and slide the body of the canister out For Op mal Performance The suc on device is to remain p...

Страница 7: ...n front cover Pa ent tubing Vinyl tubing 9 32 ID x 36 L 7 1mm ID x 91 44 cm L SSCOR part 43203 Suc on Tip HI D Big S ck SSCOR part 44241C SSCOR S3 SSCOR Part 44305C PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION Does not func on when switch is turned on DC Power Ba ery discharged Molex connec ons disconnected Damaged PC Board Reconnect to charging source to ac vate charger or install a new fully charged...

Страница 8: ...ce C Ba ery 12V DC nominal voltage D Ba ery door E Pump connec on to PC Board F Vacuum Barb and O Ring To Replace Ba ery Open ba ery door and remove the depleted ba ery Install a new ba ery as shown Make sure the contacts on the ba ery mate to the terminals on the chassis Close the ba ery door and turn the unit on to verify performance Quickdraw Model 2400 Series 8 Clearing The Airway Is Our 1 Pri...

Страница 9: ...e use not permi ed Use new canister for each pa ent Pa ent Tubing Disposable item re use not permi ed Use new pa ent tubing for each pa ent HI D S ck Disposable item re use not permi ed Use new HI D S ck for each pa ent Vacuum Pump Wipe with damp cloth or disinfectant wipe Steriliza on not permi ed Vacuum pump should be replaced if contaminated PC Board Steriliza on not permi ed PC Board should be...

Отзывы: