CHARACTERISTICS
SPECIFICATIONS
Size 12
‐
1/2"L x 4"H x 4"W (32cm L x 11cm H x 11cm W)
Weight
2.6 lbs (1.18 kg)
Vacuum Pump
12V DC, 0.7 A. Exceeds 500mmHg.
Lower levels of nega ve pressure will be observed at al tude.
Regulator
Reduces vacuum from
‐
500+mmHg (High Vacuum / Low Flow) to approximately
‐
85mmHg (Low Vacuum / Low Flow). Typically
‐
70mmHg to
‐
100mmHg but the range
can be larger due to condi ons such as variances in ba ery charge and condi on.
Ba ery
12V DC
Sealed Lead Acid replacement part #80611
‐
100
Alkaline ba ery replacement part #80613
‐
100
Please read page 6 for ba ery care sugges ons
Collec on Canister
(13)
300cc/ml capacity. Non Sterile, Single Use
Model 2480: Standard replacement Quickdraw canister (shown on page 6)
Model 2488: Barbed Quickdraw canister (shown on front cover)
Pa ent tubing
Vinyl tubing 9/32”ID x 36”L (7.1mm ID x 91.44 cm L)
SSCOR part #43203
Suc on Tip
HI
‐
D® “Big S ck®” SSCOR part #44241C
SSCOR S3 SSCOR Part #44305C
PROBLEM
POSSIBLE CAUSE
CORRECTIVE ACTION
Does not func on
when switch is
turned on (DC Power)
Ba ery discharged
Molex connec ons disconnected
Damaged PC Board
Reconnect to charging source to ac vate charger
or install a new fully charged ba ery
Open unit and re
‐
connect Molex connec ons
Replace PC Board
Ba ery Condi on
indicator lights not lit
Power cord disconnected
Damaged PC Board
Reconnect cord
Replace PC Board
No suc on when
pump is running
Canister not secure to chassis
Regulator vent is open
Tip cover is on the catheter
Fluids have shut down the
fi
lter in
the canister
Latch canister to chassis
Replace cap on vent
Remove p cover from catheter
Replace canister
Quickdraw - Model 2400 Series
7
Clearing The Airway Is Our #1 Priority
General Specifications
Trouble Shooting
Rev. M (02/17)