CHARACTERISTICS SPECIFICATIONS
Size
17”L x 9”H x 6“W
(43.2cm L x 22.9cm H x 15.2cm W)
Weight
8.5 lbs. (4.1 Kgs.)
Vacuum Pump
12V DC oil-less diaphragm. 3.0A
Exceeds -525mmHg
Clinical Airflow >= 30LPM
Variable Regulator (7)
Controls vacuum from < -50 to >= -
525mmHg
Gauge (8)
Calibrated in mmHg. Color-coded.
Power Source: Battery (DC Power)
Rechargeable Sealed Lead Acid.
45 (±10%) minute running time.
Switch (5)
On/Off rocker switch
Collection Canister (13)
1200cc/ml SSCOR part #48041
Patient tubing
Vinyl tubing 9/32”ID, 72”L
SSCOR part #43200
Suction Tip
HI-D
®
“Big Stick
®”
SSCOR part #44241
Trouble Shooting
PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION
Does not function when
Battery discharged
Reconnect to battery
switch is on
charger
Power source connected
Fuse blown
Refer to Bio-Med
indicator light not lit
Bad connection
Refer to Bio-Med
No suction when pump is
Vacuum line (10) loose
Tighten
running
Canister(13) damaged
Replace
Canister lid (13) loose
Tighten
Catheter thumb vent (6) is
open
Occlude
Debris has collected in
Refer to maintenance
pump
section of this manual
– 5 –