2 – 8
Remote Operation
2.4.4
Configuration Commands
Upper Limit
ULIM(?)
{
f
}
Set (query) the upper limit voltage
V
UL
{
to
f
}
, in volts.
If
ULIM
is set outside the range
+
10 V
≥
V
UL
≥
V
LL
+
100 mV
,
then the command will fail, setting the EXE bit in the
*ESR
register,
and setting the
LEXE
parameter to “invalid parameter” (16).
ULIM 3.14
Example:
ULIM?
+3.14
Lower Limit
LLIM(?)
{
f
}
Set (query) the lower limit voltage
V
LL
{
to
f
}
, in volts.
If
LLIM
is set outside the range
V
UL
−
100 mV
≥
V
LL
≥ −
10 V
,
then the command will fail, setting the EXE bit in the
*ESR
register,
and setting the
LEXE
parameter to “invalid parameter” (16).
LLIM -8.042
Example:
LLIM?
-8.04
Keep Clock Awake
AWAK(?)
{
z
}
Set (query) the SIM964 keep-awake mode
{
to
z
=
(
OFF 0
,
ON 1
)
}
.
Ordinarily, the clock oscillator for the SIM964 microcontroller is held
in a stopped state, and only enabled during processing of events
(Section 1.3). Setting
AWAK
ON
forces the clock to stay running, and
is useful only for diagnostic purposes.
AWAK ON
Example:
SIM964
Analog Limiter