Initial System Setup 

When  initially  set  up,  calibration  is  factory  set  and  re-calibration  is  not  needed.  The  Local 
Gravitational  Acceleration  may  have  to  be  re-set  for  the  current  geographical  location  for 
improved accuracy. Automatic Shut  Off is  shipped with  a factory default of 60 seconds.  If a 
longer  period  of  time  (up  to  900  seconds)  is  desired,  then  it  will  need  to  be  re-set.  Both 
procedures are found below. 

 

Access System Setup

  

To access all settings below, break the Calibration Seal. Press the 

calibration button 

(Figure 2) using 

the pointed-end of a pen or similar item. 

CAL mode is now active.

 

Refer to  Figure 3. The Right  Button  allows  you  to  toggle through the menus without saving any 
parameters. The Left Button permits a menu item to be selected or a parameter saved. 

NOTE

:  Ensure  that  nothing  is  in  contact  with  the  scale  system  during  this  procedure. 

Remove hands from the system when noting the displayed calibration results. 

Cal Menu 

“data”  

Displays saved calibration and model information. 

“unit”  

Set the available units; lb only, kg only, or selection of lb or kg with mode button.  

“sec”  

Set Auto Power Down Timer between 15 and 900 seconds. 

“grav”  

Set Gravitational Acceleration in m/s

2

 

“Full” 

Applied calibrated mass at full scale. 

“Half” 

Used with three point calibration curve. Applied calibrated mass at a second point 
between zero and full scale, usually at half.  

“zero” 

No calibrated weight applied. 

“2 pt” 

Two- point calibration curve, zero and full scale. 

“3 pt” 

Three-point calibration curve; zero, half and full scale. 

“save” 

Saves the calibration data. 

“quit” 

Quits without saving the calibration data. 

 

Enter CAL Mode 

To enter the CAL mode, press

 the 

calibration button

 (Figure 2) using the pointed-end of a 

pen or similar item. Toggle through the menu with Right Button (Figure 3). Select or Save 
data with the Left Button (Figure 3).

 

 

Содержание SRV411i

Страница 1: ..._171114 Page 1 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2017 SInstruments Inc S by S Instruments Inc Operating and Service Manual SRV411i Small Animal Floor Scale ...

Страница 2: ...ruments Inc TABLE OF CONTENTS TABLE OF FIGURES 2 PACKING CHECKLIST MODEL SRV411i 3 ASSEMBLY 3 SYSTEM DESCRIPTION AND INTENDED USE 4 MAINTENANCE AND CLEANING 5 STORAGE AND TRANSPORTATION 5 SPECIFICATIONS 6 BUTTON FUNCTIONS 7 BASIC SYSTEM OPERATION 8 THEORY OF OPERATION 9 CALIBRATION 10 TROUBLESHOOTING 14 DISPLAY MOUNTING TEMPLATE 15 TABLE OF FIGURES Figure 1 SRV411i Button Label 7 Figure 2 Calibrat...

Страница 3: ... by qualified personnel The mounting hardware provided will cover most wall applications B Determine a position on the wall comfortable to the user Mark the location of the two 2 mounting holes on the wall holes are spaced 2 5 on center apart using the template provided on page 15 C Drill the two 2 pilot holes no larger than diameter in the locations marked Install the two 2 self drilling screw an...

Страница 4: ...processor and load cell technology to provide accurate and repeatable weight data Four 4 identically matched transducers are strategically placed to ensure an accurate representation of the small animal s weight The small animal s weight is displayed on a LCD screen With a push of a button weight data may be viewed in either pounds or kilograms INTENDED USE The SRV411i Small Animal Floor Scale is ...

Страница 5: ...of clear polyester and can be scratched by abrasive cleaners Mild soap and water is recommended for general cleaning and disinfecting STORAGE and TRANSPORTATION To maintain proper operation of this instrument storage and transport conditions should not vary outside the following conditions Relative Humidity 0 to 85 Ambient Temperature 14 F to 122 F 10 C to 50 C DO NOT use pressurized water or stea...

Страница 6: ...ution for calibrated range AUTO ZERO One button operation AUTO WEIGH Scale will automatically turn on and lock in on a stable weight AUTO POWER DOWN After 1 minute AVERAGING Automatic digital filter POWER SUPPLY AC Wall Power Supply UL Recognized CALIBRATION Calibration is traceable to NIST standards OPERATING CONDITIONS Normal operating conditions for this product Ambient Temperature Range 68 F t...

Страница 7: ...isplayed NOTE When system is zeroed the Hold Recall memory is erased LB KG MODE BUTTON Press lb kg button to switch weight display between pounds or kilograms HOLD RECALL MODE BUTTON Scale will automatically lock a stable weight that is greater than five pounds and MEM will be displayed in the upper left of the display To manually hold a weight reading at any time during the weighing process press...

Страница 8: ...s the Zero Weigh button and wait for the display to indicate ZERO Release the Zero Weigh button and the display will read 0 0 Position the small animal on the platform Weight will be displayed in a few seconds A stable weight will automatically be stored in memory STEP 3 To manually hold a weight reading at any time during the weighing process press the Hold Recall button NOTE When the scale is of...

Страница 9: ...e operational amplifier The output of the operational amplifier is digitized by the analog to digital converter The converter integrates the analog signal onto the integrating capacitor over a short interval The integrating capacitor is then discharged at a rate proportional to the reference voltage applied to the converter The residual voltage on the integrating capacitor is then multiplied by a ...

Страница 10: ...tors on the printed circuit boards may be damaged by electrostatic discharge ESD Be sure to use proper handling precautions at all times IMPORTANT CALIBRATION CHECK Qualified service personnel only should perform this procedure Load cells have no user serviceable components and should not be tampered with for any reason Re calibration is generally not required but should be verified periodically t...

Страница 11: ...thout saving any parameters The Left Button permits a menu item to be selected or a parameter saved NOTE Ensure that nothing is in contact with the scale system during this procedure Remove hands from the system when noting the displayed calibration results Cal Menu data Displays saved calibration and model information unit Set the available units lb only kg only or selection of lb or kg with mode...

Страница 12: ...alibrated a Select the Gravitational Acceleration from the CAL menu by pressing the Left Button when the display shows grav b The display will show only the four digits to the right of the decimal point 9 XXXX c Set the local gravitational acceleration in m s2 by selecting the first digit and toggle through the digits using the Right Button until desired value is shown Press the Left Button and co...

Страница 13: ...Button again to save and the display will show SAV d Calibrate using the Three Point Calibration Curve zero half scale and full scale NOTE The scale can also be calibrated with a Two point slope 2Pt although it will not be as accurate a Select Three point calibration curve by pressing the Left Button when the display shows 3Pt b Apply no load and ensure stability Zero the scale by pressing the Lef...

Страница 14: ...N Weight reading is much lower than expected Check that the platform is clean underneath stand on each of the four corners to see if one corner is not weighing correctly System fails to perform correctly Check power supply connections Make sure the scale is clear of any obstacles and that nothing including the vet is touching the small animal or scale Set scale on smooth hard floor surface For add...

Страница 15: ... on a desk or end table If desired provisions for wall mount have been incorporated and can be accomplished in the manner described below A The display should be installed by qualified personnel The mounting hardware provided will cover most wall applications B Determine a position on the wall comfortable to the user Mark the location of the two 2 mounting holes on the wall holes are spaced 2 5 on...

Страница 16: ... MANSRV411i_171114 Page 16 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2017 SInstruments Inc S By SInstruments Inc Precision Technology in Perfect Balance ...

Отзывы: