Model SR735
i
-KG Wheelchair Scale
Operating and Service Manual
Part No. MAN735
i
-KG_MICROBAN_201208
Page 16 of 20
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
www.srscales.com
email:
Copyright 2020
S
Instruments, Inc.
CONFIGURING SEND BUTTON
STEP 1:
Remove the three (3) screws on the left end cover of the display housing. Remove the
cover. Put the scale system into the Calibration Mode by switching the calibration switch on the
display board (Figure 9).
STEP 2:
Press the “
MENU
” key until “
DATA OUT =PRINT
” is displayed
STEP 3:
Press “
UP
” or “
DOWN
” button to select between “
DATA OUT
=NONE
”, “
DATA
OUT
=
” or “
DATA OUT =EHR
”
STEP 4:
Press the “
ENTER
” button to save.
STEP 5:
Switch the scale system out of the Calibration Mode on the display board (Figure 9).
STEP 6:
Place the left end cover on the display housing. Attach using three (3) cover screws.
EHR OUTPUT
Electronic Medical Records (EMR) / Electronic Health Records (EHR) technology is supported by
the i-series scales from SR Instruments. The EMR/ EHR software is not included with scale.
There are many different EMR/ EHR software available and the connectivity of the scale to the
software requires the services of a professional.
A null modem cable is required to communicate with the scale.
Scale output when send button is configured to Electronic Health Records is as follows:
<esc>R<esc>Wnnn.n<esc>Hmmm.m<esc>Bkk.k<esc>Uuu<esc>E
Where
:
R is read
W is weight
nnn.n is the weight in kg
H is height
mmm.m is the height in cm.
B is BMI
kk.k is the BMI
U is units (kg)
uu is KG
E is end of packet.