Model SR416i-KG - Patient Floor Scale 
Operating and Service Manual 
Part No. MAN416I-KG_200611 

 

Page 5 of 16 

 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel: 716-693-5977     Fax: 716-693-5854    URL: 

www.srscales.com

 

email: 

[email protected]

     

  

Copyright 2020

 

S

Instruments, Inc. 

CLEANING and DISINFECTING 

CLEANING 

 

To clean the display / user interface and other scale contact areas: 

 

Use a soft cloth dampened with water and mild detergent to clean scale surfaces. 

 

Wipe surface with clean soft cloth dampened with water and then dry with clean soft cloth.  

 

Do  not  use  abrasive  materials  to  clean  scale  surface  to  prevent  damage  to  the  surface 
finish. 

 

Do not spray liquid directly onto scale surfaces.  Use only a damp cloth. 

 

DISINFECTION 

To disinfect the display / user interface and other scale contact areas:  

 

Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth. Cloth 
cannot  be  dripping  wet.  Follow  manufacturer’s  instruction  on  the  proper  use  of 
commercially available disinfectants. 

 

Disinfectant solutions with 1% sodium hypochlorite or 70% isopropyl alcohol are suitable 
for display / user interface and other scale contact surfaces. 

 

After  disinfecting,  use  a  soft  cloth  dampened  with  clean  water  and  dry  with  a  soft  clean 
cloth to prevent buildup of material on scale finish. 

 

Do  not  use  abrasive  material  to  disinfect  /  clean  scale  surfaces  to  prevent  damage  to  the 
surface finish. 

 

Do not spray liquid directly onto scale surfaces. Use only a damp cloth. 

 

WARNING:

 

DO  NOT  SPRAY  CLEANING  SOLUTION  OR  LIQUIDS  DIRECTLY 
ON SURFACES TO BE CLEANED

 

WARNING:  EXPOSURE  TO  EXCESSIVE  LIQUID  WILL  DAMAGE  USER 

INTERFACE KEYPAD. 

 

WARNING:

 

DO  NOT 

USE  PRESSURIZED  WATER  OR  STEAM.  THE  SCALE 

SYSTEM CONTAINS ELECTRONIC COMPONENTS THAT MAY BE 

ADVERSELY  AFFECTED  BY  EXPOSURE  TO  SUCH  AN 

ENVIRONMENT. 

Содержание SR416i-KG

Страница 1: ...KG_200611 Page 1 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S I...

Страница 2: ...struments Inc TABLE OF CONTENTS PACKING CHECKLIST MODEL SR416I KG 3 ASSEMBLY 3 SYSTEM DESCRIPTION AND INTENDED USE 3 STORAGE AND TRANSPORTATION 4 MICROBAN ANTIMICROBIAL PROTECTION 4 CLEANING AND DISIN...

Страница 3: ...power adapter into a wall outlet SYSTEM DESCRIPTION and INTENDED USE SYSTEM DESCRIPTION The SR416i KG Patient Floor Scale employs the latest in microprocessor and load cell technology to provide accu...

Страница 4: ...TATION To maintain proper operation of this instrument storage and transport conditions should not vary outside the following conditions Relative Humidity Range of 0 to 85 Ambient Temperature Range of...

Страница 5: ...eas Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disin...

Страница 6: ...ibrated range AUTO ZERO One button operation AUTO WEIGH Scale will automatically turn on and lock in on a stable weight AUTO POWER DOWN After 1 minute AVERAGING Automatic digital filter POWER SUPPLY M...

Страница 7: ...play reads ZERO then release and wait until the display reads 0 0 When the patient is in position on the scale the weight will be displayed NOTE When the system is zeroed the Hold Recall memory is era...

Страница 8: ...rnatively press the Zero Weigh button and wait for the display to indicate ZERO Release the Zero Weigh button and the display will read 0 0 Position the patient on the platform Weight will be displaye...

Страница 9: ...by the operational amplifier The output of the operational amplifier is digitized by the analog to digital converter The converter integrates the analog signal onto the integrating capacitor over a sh...

Страница 10: ...et in calibration mode Both procedures are found below To access all settings below remove the feet spacers and seven 7 screws holding the bottom cover on Press the calibration button Figure 2 and CAL...

Страница 11: ...n of lb or kg with mode button sec Set Auto Power Down Timer between 15 and 900 seconds grav Set Gravitational Acceleration in m s2 Full Applied calibrated mass at full scale Half Used with three poin...

Страница 12: ...cceleration from the CAL menu by pressing the Left Button when the display shows grav b The display will show only the four digits to the right of the decimal point 9 XXXX c Set the local gravitationa...

Страница 13: ...t by pressing the Left Button to quit without saving changes e Press the Left Button again to save and the display will show SAV d Calibrate using the Two Point Calibration Curve zero and full scale N...

Страница 14: ...ECTIVE ACTION Weight reading is much lower than expected Check that the platform is clear underneath stand on each corner to see if one corner is not weighing correctly System fails to perform correct...

Страница 15: ...liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In n...

Страница 16: ...rt No MAN416I KG_200611 Page 16 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstrument...

Отзывы: