CALIBRATION 

 

STEP 1: 

Select weights traceable to NIST, up to the Full Scale value (maximum capacity). 

STEP 2:

  Remove the center Top Panel and disconnect the two (2) cables from the Transducers. Turn 

the  Top  Panel  over  and  remove  the  four  (4)  screws  and  the  clear  Plastic  Cover  from  the 
Display housing. 

STEP 3

Remove  the  four  (4)  screws  holding  the  Display  housing  to  the  Top  Panel  and  remove 
housing. Run the two cables from the Display housing under the Transducer and reconnect 
to the Transducer connectors. 

STEP 4:

  Slide  the  Top Panel  back  in  place over the Transducer  assembly  pins. Ensure the  Display 

housing is out from under the scale assembly. 

STEP 5:

  Press the calibration button “

S2

” (Figure 8). The display will read “

CAL

”. 

STEP 6:

  Press button “

S4

” to scroll through menu options until “

FULL

” is displayed. Press the “

S1

” 

button to select. 

STEP 7:

  Set the “

FULL

” value to the selected weight from Step 1. Use the “

S1

” button to select digit 

positions and use button “

S4

” to change the value. 

STEP 8:

  When finished, the display will read “

SAVE

”. Press the “

S1

” button to save, or press button 

S4

” then “

S1

” to “

QUIT

”. 

STEP 9:

  Press “

S4

” to scroll to the menu option “

2 PT

” and press the “

S1

” button. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

IMPORTANT 

CALIBRATION  CHECK  Qualified  service  personnel  only  should  perform  this 
procedure. Load cells have no user serviceable components and should not be tampered 
with  for  any  reason.  Re-calibration  is  generally  not  required,  but  should  be  verified 
periodically  to  ensure  accuracy.  The  recommendation  for  calibration  check  is  at  least 
once every 12 months, or as individual maintenance policy requires. 

 

Continued on next page 

Figure 8: Calibration Buttons 

S1 

S4 

CAL 

Содержание SRScales SRV714 Series

Страница 1: ...age 1 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S Instruments I...

Страница 2: ...ARTS and ACCESSORIES 6 SYSTEM DESCRIPTION and INTENDED USE 6 MAINTENANCE and CLEANING 7 STORAGE and TRANSPORTATION 7 THEORY OF OPERATION 7 SPECIFICATIONS 8 BASIC SYSTEM OPERATION 9 BATTERY REPLACEMENT...

Страница 3: ...ruments com Copyright 2020 SInstruments Inc PACKING CHECKLIST Model SRV714 Large Animal Platform Scale DESCRIPTION QUANTITY TOP PANELS 47 in 119 cm x 31 75 in 81 cm 3 ea TRANSDUCERS 95 3 8 in 242 cm x...

Страница 4: ...to battery holder imprinted instructions Attach the battery holder cover and tighten the thumb screws Re secure the plastic cover with the four 4 screws STEP 3 Remove the other two end panels from th...

Страница 5: ...les are in line with the connectors on the transducer tubes Figure 4 STEP 8 After the transducers are plugged in carefully align the center panel with the four 4 alignment pins that are located in the...

Страница 6: ...Scale derives its power from six 6 each D cell batteries Weight is displayed with a displayed resolution of 5 pounds or 2 kilograms The SRV714 Large Animal Platform Scale is equipped with a Bluetooth...

Страница 7: ...to a digital signal by an analog to digital converter The digital signal is transferred to a micro controller where it is filtered converted to appropriate units and displayed on a liquid crystal dis...

Страница 8: ...BLUETOOTH ON TIME 10 minutes from last button press AVERAGING Automatic digital filter POWER SUPPLY Six 6 D cell batteries CALIBRATION Calibration is traceable to NIST standards OPERATING CONDITIONS N...

Страница 9: ...ots in the upper right corner of the SR App and select to Connect a device Figure 6B STEP 5 Select the SRV714 Bluetooth module from list of devices Figure 6C If the device is not listed the display ma...

Страница 10: ...ry replacement is needed an indicator will appear on the display STEP 1 Remove the center Top Panel Figure 4 and disconnect the two 2 cables from the Transducers Turn the Top Panel over and remove the...

Страница 11: ...the Display housing is out from under the scale assembly STEP 5 Press the calibration button S2 Figure 8 The display will read CAL STEP 6 Press button S4 to scroll through menu options until FULL is d...

Страница 12: ...Display to the platform and securely tighten screws Re attach plastic cover to Display NOTE Tighten screws only until rubber seal starts to compress STEP 15 Connect the two Display cables to Transduc...

Страница 13: ...on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event sha...

Страница 14: ...AN714_200601 Page 14 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S By...

Отзывы: