SR Instruments SR725 Скачать руководство пользователя страница 10

Model SR725 Series Wheelchair Scale System 
Operating and Service Manual - S/N 4068+ 
Part No. MAN725_100928 

 

Page 10 of 16 

 
 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel:  716-693-5977     Fax:  716-693-5854    URL:  

www.srscales.com

 

email:  

[email protected]

     Copyright

 

2010

 

S

Instruments, Inc. 

BASIC SYSTEM OPERATION cont’d

 

STEP 1

:  Place the empty wheelchair (with any blankets or pillows that the patient may have with 

them) on the scale platform.  Then press the zero button.   

NOTE

:  The display will shut off after 35 seconds. The zero (TARE) will still be stored in 

memory.   

STEP 2

:  Remove the wheelchair from the platform.  Place the patient in the wheelchair.  

STEP 3

:  Wheel the patient with wheelchair onto the platform. Lock the patient’s wheelchair brake to 

prevent movement.  Press the weigh button.  The display will show the patient’s weight only.   

 
 
 
 
 
CONTINUOUS WEIGH MODE   

In this default mode, the weighing surface remains active.  Press the “

HOLD

” button once to lock the 

displayed reading and store it in memory as the “last weight” for recall later if needed. 

AUTO-HOLD MODE    

This mode is for patients unable to remain still for the weighing procedure.  It locks, stores, and displays 
the patient’s weight as soon as the “

WEIGH

” button is pressed once.  

Note

:  No weight will be 

displayed until the button is pressed.   

To enable this mode before zeroing the system, press and hold the “

HOLD

” button for approximately 

five (5) seconds until the display reads “

AUTO-HOLD ENABLED

”. 

To return to CONTINUOUS WEIGH mode when finished, press and hold the “

HOLD

” button for 

approximately five (5) seconds until the display reads “

CONTINUOUS WEIGH

”. 

 

BATTERY REPLACEMENT 

STEP 1

:  The display will read “

REPLACE BATTERY

”.  

STEP 2

:  (Figure 8) Unscrew the panel screw on the Battery Compartment Cover (8) and remove the 

battery compartment cover.   

STEP 3

:  Remove and replace ALL six (6) “D” cell batteries.  Refer to diagram in the battery 

compartment for placement.   

STEP 4

: Press 

the 

WEIGH

” button to confirm display is working. 

STEP 5

:  Secure the battery cover using the panel screw.    

STEP 6

:  Zero the system.  

 

 

 

 

Figure 8:    Battery Compartment Cover Panel Screw 

8

CAUTION 

 

Do not leave patient unattended on the scale platform. 

Содержание SR725

Страница 1: ...0928 Page 1 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2010 SInstruments Inc by S Instrumen...

Страница 2: ...ESCRIPTION AND INTENDED USE 6 MAINTENANCE AND CLEANING 7 STORAGE AND TRANSPORTATION 7 SPECIFICATIONS 8 BUTTON FUNCTIONS 9 BASIC SYSTEM OPERATION 9 BATTERY REPLACEMENT 10 THEORY OF OPERATION 11 CALIBRA...

Страница 3: ...Wheelchair Scale System PACKING CHECKLIST SR725L Wheelchair Scale System DESCRIPTION QUANTITY BASE ASSEMBLY 32 in x 32 in 81 cm x 81 cm WITH ATTACHED MAST PIPE AND RAMP 1 ea DISPLAY UNIT 1 ea PACKAGE...

Страница 4: ...ipe 2 on the floor STEP 4 Loosen the Mast Wing Nut 4 to unlock the Mast Pipe Pull the Mast Pipe out and away from the scale to free it from the internal locking pin and swing it up until the Mast Pipe...

Страница 5: ...over 6 Install the six 6 D cell batteries as indicated on the plastic battery cradle Replace the cover Figure 4 STEP 9 Adjust leveling feet located in each of the four corners of the underside of the...

Страница 6: ...power microprocessor circuitry allows the SR725 Wheelchair Scale System to derive its power from six 6 common D cell batteries that will provide up to 10 000 weight readings before needing replacement...

Страница 7: ...ing conditions Relative Humidity 0 to 85 Ambient Temperature 14 F to 122 F 10 C to 50 C TRANSPORTATION To transport the scale fold Ramp see Figure 4 into the closed position and lock with the Ramp Loc...

Страница 8: ...ated range AUTO ZERO One button operation AUTO POWER DOWN Approximately 35 seconds HOLD Stores displayed reading in memory LAST WEIGHT RECALL Press HOLD button to recall last stored displayed weight A...

Страница 9: ...ould Auto Power Down before the weighing process is done HOLD The HOLD button freezes the displayed weight and stores it away in memory Press HOLD to store the weight into memory To recall last weight...

Страница 10: ...and store it in memory as the last weight for recall later if needed AUTO HOLD MODE This mode is for patients unable to remain still for the weighing procedure It locks stores and displays the patient...

Страница 11: ...the analog to digital converter The converter integrates the analog signal onto the integrating capacitor over a short interval The integrating capacitor is then discharged at a rate proportional to...

Страница 12: ...en in the CAL mode press the ZERO button to zero the display STEP 5 Place a known calibrated weight traceable to NIST onto the weighing surface and compare it to the displayed reading Note DO NOT USE...

Страница 13: ...SW5 are located internally on the PC Board Figure 10 STEP 2 Simultaneously press buttons indicated to initialize the system The display will read HOLD TO INIT and count down from 9 to 0 When the init...

Страница 14: ...tery Wait five seconds then re install the battery and try the WEIGH button again The display lights appear to work but do not respond to button activation Button may not have bounced back up after be...

Страница 15: ...ments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instrum...

Страница 16: ...S N 4068 Part No MAN725_100928 Page 16 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2010 SIns...

Отзывы: