CALIBRATION 

 
NOTE:

  Ensure that nothing is in contact with the scale system during this procedure.  Remove 

hands from the system when noting the displayed calibration results. 

 
CHECKING CALIBRATION 
 
STEP 1:

  Select known calibrated weights, traceable to NIST.   

NOTE: DO NOT USE

 barbells or un-calibrated weights. 

STEP 2:

  Zero the scale by pressing the 

ZERO 

button. 

STEP 3:

  Place the calibrated weight on the scale.  Wait for scale to 

stabilize; note scale reading. 

STEP 4:

  Scale readings should be within Calibration  

Tolerance Table 

(

Figure 4

). 

 

 

SETTING CALIBRATION 
 
 
 
 
 
 
 
 

NOTE:  

Ensure that nothing is in contact with the scale during this procedure and that all six (6) 

feet  are  touching  a  level  surface.    Remove  hands  from  the  system  when  noting  the  displayed 
calibration results.  (Use known calibrated weight traceable to NIST)

 

STEP 1:

  Enter the System Settings menu and select “Calibrate Scale”. 

STEP 2:

  Remove all weight from the platform and press “

Done

”. 

STEP 3:

  Place the calibrated weight on the scale.  Wait for scale to stabilize. 

STEP 4:

  Press the “WEIGHT” field and enter the value of the weight on the platform. 

STEP 5: 

 Press the green checkmark        then press the “Save” button. 

STEP 6: 

 Display will automatically go to Home page with displayed weight. Remove weight, 

zero scale and then place weight on scale to verify calibration is correct. 

 

CALIBRATION 

TOLERANCE TABLE 

LOW 

LIMIT 

APPLIED 

LOAD 

HIGH 

LIMIT 

99.8 

100.0 

100.2 

199.6 

200.0 

200.4 

299.4 

300.0 

300.6 

399.2 

400.0 

400.8 

Figure 4:  Calibration Table 

IMPORTANT 

CALIBRATION  CHECK      Qualified  service  personnel  only  should  perform  this 
procedure.  Load cells have no user serviceable components and should not be tampered 
with  for  any  reason.    Re-calibration  is  generally  not  required,  but  should  be  verified 
periodically  to  ensure  accuracy.    The  recommendation  for  calibration  check  is  at  least 
once every 12 months, or as individual maintenance policy requires. 

 

CAUTION 

The  integrated  circuits  and  semiconductors  on 
the  printed  circuit  boards  may  be  damaged 
by  electrostatic  discharge  (ESD).    Be  sure  to 
use proper handling precautions at all times.  

Содержание SR Scales SRV95 Series

Страница 1: ...AN955_220721 Page 1 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2022 S Instruments Inc SRV955 WALL MOUNT DISPLAY PLATFORM SCALE S by SR Instruments Inc Operating and Service Manual ...

Страница 2: ...ts Inc TABLE OF CONTENTS TABLE OF FIGURES 2 PACKING CHECKLIST MODEL SRV955 3 ASSEMBLY 3 SYSTEM DESCRIPTION AND INTENDED USE 4 STORAGE AND TRANSPORTATION 6 CLEANING AND DISINFECTING 5 SPECIFICATIONS 6 DISPLAY FUNCTIONS 7 BASIC SYSTEM OPERATION 7 SYSTEM SETTINGS 8 REMOTE DISPLAY 10 CALIBRATION 10 TROUBLESHOOTING 11 WARRANTY 13 DISPLAY MOUNTING TEMPLATE 15 TABLE OF FIGURES Figure 1 Display Screen 7 F...

Страница 3: ...ll cover most wall applications B Determine a position on the wall comfortable to the user Mark the location of the two 2 mounting holes on the wall holes are spaced 2 5 on center apart using the template provided on page 15 C Drill the two 2 pilot holes no larger than diameter in the locations marked Install the two 2 self drilling screw anchors D Install the two 2 8 x 1 flat head screws supplied...

Страница 4: ...ology to provide accurate and repeatable weight data Four 4 identically matched transducers are strategically placed to ensure an accurate representation of the animal s weight The animal s weight is displayed on an LCD screen and can be viewed in either pounds or kilograms INTENDED USE The SRV955 Portable Platform Scale is a lightweight portable scale designed for use as a weighing system for ani...

Страница 5: ...ION Follow manufacturer s instruction on the proper use of commercially available disinfectants Disinfectant solutions with 1 sodium hypochlorite or 70 isopropyl alcohol are suitable for display user interface and other scale contact surfaces To disinfect the display user interface Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Afte...

Страница 6: ...ility SPECIFICATIONS MAXIMUM WEIGHT CAPACITY 400 lb 182 kg PLATFORM SIZE 24 in x 36 in x 2 in 61 cm x 92 cm x 5 cm DISPLAY TYPE 4 3 TFT Color LCD Resistive Touch DISPLAY RESOLUTION 0 1 lb 0 1 kg ACCURACY 0 2 2 digits of displayed resolution for calibrated range AUTO ZERO One touch operation AVERAGING Automatic digital filter POWER SUPPLY AC Wall Power Supply Medical Grade CALIBRATION Calibration i...

Страница 7: ...weigh mode and when there is a weight value stored The ZERO button is used to zero the system before placing a patient onto the scale system and to erase the last saved weight Ensure that nothing is in contact with the weighting surface during the procedure The display will read 0 0 LB or KG The Stable Weight icon blinks while actively weighing and then becomes steady when weight is stable The Set...

Страница 8: ...t will be displayed until the display is zeroed again or until the RECALL button is pressed again NOTE The RECALL button is not displayed when there is no stored weight or when in Continuous weigh mode SLEEP MODE With no weight on the platform or after the display is zeroed the display will power down to a sleep mode after 20 seconds Touch the screen to wake it up CONTINUOUS WEIGH With Continuous ...

Страница 9: ...e screen press the System Settings button STEP 2 Enter password 12345 STEP 3 Press the green checkmark SETTINGS Weight Data Weight units can be set to LB or KG Display can be set to Auto Hold or Continuous weigh modes Remote Display Main display can be paired to a remote display sold separately See Remote Display section for pairing instructions Calibrate Scale Recalibrate Scale platform System In...

Страница 10: ...liary Display was purchased separately follow the instructions below TO PAIR A REMOTE DISPLAY STEP 1 Standing near Primary display connect power cord to remote display STEP 2 Access System Setting Menu and select Remote Display STEP 3 Press the PAIR button on Wireless Auxiliary Display STEP 4 Verify that Wi Fi icon on Wireless Auxiliary Display is illuminated green NOTE A WIRELESS AUXILIARY DISPLA...

Страница 11: ...aceable to NIST STEP 1 Enter the System Settings menu and select Calibrate Scale STEP 2 Remove all weight from the platform and press Done STEP 3 Place the calibrated weight on the scale Wait for scale to stabilize STEP 4 Press the WEIGHT field and enter the value of the weight on the platform STEP 5 Press the green checkmark then press the Save button STEP 6 Display will automatically go to Home ...

Страница 12: ...isplay reads LOW Ensure scale is on a level surface Check to make sure all four feet are touching ground and ensure that nothing is touching bottom or sides of platform Display reads 0 0 Ensure Signal Cable is connected There should be a weight difference between having the Signal Cable connected compared to being disconnected Display does not turn on Power may be lost Check circuit breaker switch...

Страница 13: ...s or liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instruments Inc be liable for any subsequent or special damages Any misuse improper installation or tampering shall void this warranty DAMAGED SHIPMENTS Title passes to purchaser...

Страница 14: ...tform Operating and Service Manual Part No MAN955_220721 Page 14 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2022 S Instruments Inc NOTES ...

Страница 15: ...r end table If desired provisions for wall mount have been incorporated and can be accomplished in the manner described below A The display should be installed by qualified personnel The mounting hardware provided will cover most wall applications B Determine a position on the wall comfortable to the user Mark the location of the two 2 mounting holes on the wall holes are spaced 2 5 on center apar...

Страница 16: ...Manual Part No MAN955_220721 Page 16 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2022 S Instruments Inc S By SInstruments Inc Precision Technology in Perfect Balance ...

Отзывы: