Model SRV947KG 24 X 48 Platform Scale
Operating and Service Manual
Part No.: MAN947KG_220429
Page 12 of 14
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
www.srscales.com
email:
Copyright 2022
S
Instruments, Inc.
CALIBRATION cont’d
STEP 8
: Display will read “ZERO”. Ensure no weight is on the platform, then press
“
HOLD/RECALL
”
STEP 9
: Display will read “FULL”. Add the selected weight to the platform. Wait for the
weight to stabilize, then press “
HOLD/RECALL
”
STEP 10
: Press “
HOLD/RECALL
” to cycle through the number “.8037”.
DO NOT CHANGE
THIS NUMBER
STEP 11
: When display reads “SAVE”, press the “
HOLD/RECALL
” button to cycle through
“Savd” and return to “CAL”
STEP 12
: Calibration is completed. Press the left button on rear of display to return to run mode.
Allow time for scale to turn off.
TROUBLESHOOTING
SYMPTOM
REASON/CORRECTIVE ACTION
Display reads LOW
Battery needs to be replaced.
Display reads LOW
Ensure scale is on a level surface. Check to make
sure all six feet are touching ground and ensure
that nothing is touching bottom or sides of
platform.
For additional information or assistance, phone our Service Hotline: 1-800-654-6360
or e-mail:
CAUTION
The integrated circuits and semiconductors on
the printed circuit boards may be damaged
by electrostatic discharge (ESD). Be sure to
use proper handling precautions at all times.