Model SRV947-BT 24 X 48 Platform Scale App-Enabled
Operating and Service Manual
Part No.: MAN947-BT_220429
Page 13 of 16
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
www.srscales.com
email:
Copyright 2022
S
Instruments, Inc.
CALIBRATION cont’d
SETTING CALIBRATION
NOTE
:
Ensure that nothing is in contact with the scale system during this procedure and that
all six (6) feet are touching a level surface. Remove hands from the system when noting the
displayed calibration results. (Use a known calibrated weight traceable to NIST.)
STEP 1
: Remove the display cover from bottom of scale
STEP 2
: Press the “
ZERO/WEIGH
” button to turn the scale on.
STEP 3
:
Press “
Lb/Kg
”
button and toggle to select lb.
STEP 4
:
Press the
Calibration
button on rear of display. Display will read “
CAL
”
STEP 5
: Toggle the “
ZERO/WEIGH
” button to read “
FULL
” then press “
HOLD/RECALL
”.
STEP 6
: Display will read “
400.0
”, set this value to the value of the selected weight
Press “
HOLD/RECALL
” to select a digit position
Press the “
ZERO/WEIGH
” button to change the value of the digit
When display reads “
SAVE
”, press the “
HOLD/RECALL
” button to cycle through
“
Savd
” and return to “
CAL
”
STEP 7
: Toggle the “
ZERO/WEIGH
” button to read “
2 Pt
” then press “
HOLD/RECALL
”.
Figure 7: Calibration Button
(Underside of Display Board)
Figure 8: Button Display
Continued next page
CAUTION
The integrated circuits and semiconductors on
the printed circuit boards may be damaged
by electrostatic discharge (ESD). Be sure to
use proper handling precautions at all times.