CALIBRATION 

 

 
STEP 1:

  Select known calibrated weights, traceable to NIST. 

NOTE: 

CALIBRATED  WEIGHT  SHOULD  BE  EQUAL  TO  OR  GREATER  THAN  THE 
MAXIMUM  PATIENT  WEIGHT,  NOT  TO  EXCEED  THE  MAXIMUM 
CAPACITY OF THE SCALE. DO NOT USE UNCALIBRATED WEIGHTS. 

STEP 2:

  Remove the two (2) screws holding the display housing to the platform 

STEP 3:

  Pull  the  display  out  from  under  the  platform  and  remove  the  eight  (8)  screws  from  the 

display box cover pictured in (Figure 4).   

STEP 4:

  Press the calibration button (Figure 5).  The display will read “

CAL

”. 

STEP 5:

   Press  button  “

S4

”  (Figure 5) to scroll through menu options until  “

FULL

” is displayed.  

Press the 

DISPLAY

 button. 

STEP 6: 

Set  the  “

FULL

” value to the selected  weight  from Step  1.  Use the 

DISPLAY

  button  to 

select digit positions and use button 

S4

 to change the value. 

STEP 7:

  When finished, the display will read “

SAVE

”.  Press the 

DISPLAY

 button to save, or press 

button 

S4 

then 

DISPLAY

 to “

QUIT

”. 

STEP 8:

  Press 

S4

 to scroll to the menu option “

2 PT

” and press the 

DISPLAY 

button. 

 

 

 

 

 

 

 

 

 

 

STEP 9:

  Ensure  nothing  is  in  contact  with  the  platform  then  press  the 

DISPLAY

  button  when  the 

display reads “

ZERO

”. 

IMPORTANT 

CALIBRATION  CHECK      Qualified  service  personnel  only  should  perform  this 
procedure.  Load cells have no user serviceable components and should not be tampered 
with  for  any  reason.    Re-calibration  is  generally  not  required,  but  should  be  verified 
periodically  to  ensure  accuracy.    The  recommendation  for  calibration  check  is  at  least 
once every 12 months, or as individual maintenance policy requires. 

 

Continued next page 

S4 

Display 

Lower Limit Applied Weight Upper Limit

99.8

100

100.2

199.6

200

200.4

249.5

250

250.5

399.2

400

400.8

499.0

500

501.0

748.5

750

751.5

998.0

1000

1002.0

1497.0

1500

1503.0

1996.0

2000

2004.0

Calibration Tolerance Chart

CAL 

Front side 
of board 

Figure 5:  Calibration Buttons 

All manuals and user guides at all-guides.com

Содержание SR Scales SRV716W

Страница 1: ...600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2021 SInstruments Inc S by SR Instruments Inc Operating and Service Manual Marine Scale System App Enabled Wireless Model SRV716W All manuals and user guides at all guides com ...

Страница 2: ...4 REPLACEMENT PARTS AND ACCESSORIES 4 SYSTEM DESCRIPTION AND INTENDED USE 5 STORAGE AND TRANSPORTATION 5 CLEANING AND DISINFECTING 6 SPECIFICATIONS 7 BUTTON FUNCTIONS 8 BASIC SYSTEM OPERATION 8 SR SCALES APP 9 BATTERY REPLACEMENT 10 THEORY OF OPERATION 11 CALIBRATION 12 TROUBLESHOOTING 13 WARRANTY 14 NOTES 15 TABLE OF FIGURES Figure 1 Battery Compartment 4 Figure 2 LB KG Mode Pushbutton Switch 8 F...

Страница 3: ...77 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2021 SInstruments Inc PACKING CHECKLIST Model SRV716W Marine Platform Scale DESCRIPTION QUANTITY PLATFORM SCALE 1 ea PACK OF SIX 6 AA BATTERIES 1 ea CALIBRATION CERTIFICATE 1 ea WARRANTY CARD 1 ea TABLET AND CASE 1 ea MANUAL 1 ea All manuals and user guides at all guides com ...

Страница 4: ... parts call the Service Hotline at 1 800 654 6360 STEP 2 Place the Platform upside down on the floor and remove the cover on the Battery Compartment STEP 3 Install six 6 AA cell batteries and select desired units of weight to be displayed with LB KG switch according to BATTERY REPLACEMENT instructions STEP 4 Turn platform right side up and press the DISPLAY button to confirm display is working Fig...

Страница 5: ...olution of 0 2 pounds or 0 1 kilograms The SRV716W Marine Platform Scale is equipped with a Bluetooth wireless module By waking the scale up with the weigh button the wireless feature is turned on Then use the free SR Scales app Google Play store to connect with the scale to allow for remote wireless operation INTENDED USE The SRV716W Marine Platform Scale is designed for use with marine mammals p...

Страница 6: ...ith disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disinfectants Disinfectant solutions with 1 sodium hypochlorite or 70 isopropyl alcohol are suitable for display user interface and other scale contact surfaces After disinfecting use a soft cloth dampened with clean water and dry with ...

Страница 7: ...utes from last button press AVERAGING Automatic digital filter POWER SUPPLY Six 6 AA cell batteries CALIBRATION Calibration is traceable to NIST standards OPERATING CONDITIONS Normal operating conditions for this product Ambient Temperature Range 40 F to 95 F 5 C to 35 C Relative Humidity Range 0 to 85 Avoid exposure to high pressure water or steam TRANSPORTATION AND STORAGE Storage and transport ...

Страница 8: ...ct the mode for weighing with this Toggle Button BASIC SYSTEM OPERATION Ensure that the scale is free and clear of any obstructions before operating Press the DISPLAY button and hold it for 2 3 seconds The display will read 0 0 and be ready for weighing When a mat or pad is to be used on the scale platform the weight of the pad must be tared out Place the mat on the platform then press and hold th...

Страница 9: ...r Android device to display weight and control the SRV716W scale functions Figure 3 Ensure your remote display indicates CONNECTED before use The WEIGH button will display weight in pounds or kilograms as selected The Held Wt button allows access to the last stable weight that was achieved The ZERO button is used to zero the system The KILOS button allows weight data to be viewed in kilograms disp...

Страница 10: ...ndicator will appear on the display STEP 1 Turn the platform over and remove the four screws on the Battery Cover Figure 4 STEP 2 Remove and replace the six 6 AA cell batteries STEP 3 Press the DISPLAY button to confirm display is working Verify that the desired units of weight LB KG are active STEP 4 Replace the screws in the Battery Cover NOTE TIGHTEN SCREWS ONLY UNTIL RUBBER SEAL STARTS TO COMP...

Страница 11: ...nted in a full Wheatstone bridge configuration These bridges convert the physical movement of the force cell due to the applied mass on the system into minute changes in electrical resistance These changes in resistance produce a voltage difference across the Wheatstone bridge which is amplified by the operational amplifier The amplifier is configured to current sum the output of each cell with po...

Страница 12: ... the DISPLAY button to select digit positions and use button S4 to change the value STEP 7 When finished the display will read SAVE Press the DISPLAY button to save or press button S4 then DISPLAY to QUIT STEP 8 Press S4 to scroll to the menu option 2 PT and press the DISPLAY button STEP 9 Ensure nothing is in contact with the platform then press the DISPLAY button when the display reads ZERO IMPO...

Страница 13: ...NOTE TIGHTEN SCREWS ONLY UNTIL RUBBER SEAL STARTS TO COMPRESS STEP 14 Press and hold the DISPLAY button to zero the platform When the display indicates 0 0 place the calibrated weight on the scale again and check against the weight displayed If the weight is not correct recalibrate If calibration cannot be accomplished call the Service Department TROUBLESHOOTING SYMPTOM REASON CORRECTIVE ACTION We...

Страница 14: ...c s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instruments Inc be liable for any subsequent or special damages Any misuse improper installation or tampering shall void this warranty DAMAGED SHIPMENTS Title passes to purchaser upon delivery to Transportation Comp...

Страница 15: ...rt No MAN716W_210528 Page 15 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2021 SInstruments Inc NOTES All manuals and user guides at all guides com ...

Страница 16: ...of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2021 SInstruments Inc S By S Instruments Inc Precision Technology in Perfect Balance All manuals and user guides at all guides com ...

Отзывы: