Model SRV712W Marine Scale System
Service Manual
Part No. MAN712W_170816
Page 11 of 14
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
email:
Copyright 2017
S
Instruments, Inc.
CALIBRATION
STEP 1:
Select known calibrated weights, traceable to NIST.
NOTE:
CALIBRATED WEIGHT SHOULD BE EQUAL TO OR GREATER THAN THE
MAXIMUM PATIENT WEIGHT, NOT TO EXCEED THE MAXIMUM
CAPACITY OF THE SCALE. DO NOT USE UNCALIBRATED WEIGHTS
STEP 2:
Remove the two (2) screws holding the display housing to the platform
STEP 3:
Pull the display out from under the platform and remove the eight (8) screws from the
display box cover pictured in Figure 4.
STEP 4:
Press the calibration button (Figure 5). The display will read “
CAL
”.
STEP 5:
Press button “
S4
” (Figure 5) to scroll through menu options until “
FULL
” is displayed.
Press the
DISPLAY
button.
STEP 6:
Set the “
FULL
” value to the selected weight from Step 1. Use the
DISPLAY
button to
select digit positions and use button
S4
to change the value.
STEP 7:
When finished, the display will read “
SAVE
”. Press the
DISPLAY
button to save, or press
button
S4
then
DISPLAY
to “
QUIT
”.
STEP 8:
Press
S4
to scroll to the menu option “
2 PT
” and press the
DISPLAY
button.
Figure 5: Calibration Buttons
STEP 9:
Ensure nothing is in contact with the platform then press the
DISPLAY
button when the
display reads “
ZERO
”.
IMPORTANT
CALIBRATION CHECK Qualified service personnel only should perform this
procedure. Load cells have no user serviceable components and should not be tampered
with for any reason. Re-calibration is generally not required, but should be verified
periodically to ensure accuracy. The recommendation for calibration check is at least
once every 12 months, or as individual maintenance policy requires.
Continued next page
S4
CAL
Display
Lower Limit Applied Weight Upper Limit
99.8
100
100.2
199.6
200
200.4
249.5
250
250.5
399.2
400
400.8
499.0
500
501.0
748.5
750
751.5
998.0
1000
1002.0
1497.0
1500
1503.0
1996.0
2000
2004.0
Calibration Tolerance Chart