Model SRV712W Marine Scale System 
Service Manual  
Part No. MAN712W_170816 

 

Page 11 of 14 

 
 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel:  716-693-5977     Fax:  716-693-5854    URL:  

www.srscales.com

 

email:  

[email protected]

     

  

Copyright 2017

 

S

Instruments, Inc. 

 

CALIBRATION 

 

 
STEP 1:

  Select known calibrated weights, traceable to NIST. 

NOTE:  

CALIBRATED WEIGHT SHOULD BE EQUAL TO OR GREATER THAN THE 
MAXIMUM PATIENT WEIGHT, NOT TO EXCEED THE MAXIMUM 
CAPACITY OF THE SCALE. DO NOT USE UNCALIBRATED WEIGHTS 

STEP 2: 

Remove the two (2) screws holding the display housing to the platform

 

STEP 3: 

Pull the display out from under the platform and remove the eight (8) screws from the 
display box cover pictured in Figure 4.  

 

STEP 4: 

Press the calibration button (Figure 5).  The display will read “

CAL

”.

 

STEP 5: 

Press button “

S4

” (Figure 5) to scroll through menu options until “

FULL

” is displayed.  

Press the 

DISPLAY

 button.

 

STEP 6: 

Set the “

FULL

” value to the selected weight from Step 1.  Use the 

DISPLAY

 button to 

select digit positions and use button 

S4

 to change the value.

 

STEP 7: 

When finished, the display will read “

SAVE

”.  Press the 

DISPLAY

 button to save, or press 

button 

S4

 then 

DISPLAY

 to “

QUIT

”.

 

STEP 8: 

Press 

S4

 to scroll to the menu option “

2 PT

” and press the 

DISPLAY

 button.

 

 

 

 

 

 

 

 

 

 

 Figure 5: Calibration Buttons 

 

STEP 9: 

Ensure nothing is in contact with the platform then press the 

DISPLAY

 button when the 

display reads “

ZERO

”.

 

IMPORTANT 

CALIBRATION  CHECK      Qualified  service  personnel  only  should  perform  this 
procedure.  Load cells have no user serviceable components and should not be tampered 
with  for  any  reason.    Re-calibration  is  generally  not  required,  but  should  be  verified 
periodically  to  ensure  accuracy.    The  recommendation  for  calibration  check  is  at  least 
once every 12 months, or as individual maintenance policy requires. 

 

Continued next page 

S4 

CAL 

Display 

Lower Limit Applied Weight Upper Limit

99.8

100

100.2

199.6

200

200.4

249.5

250

250.5

399.2

400

400.8

499.0

500

501.0

748.5

750

751.5

998.0

1000

1002.0

1497.0

1500

1503.0

1996.0

2000

2004.0

Calibration Tolerance Chart

Содержание SR Scales SRV712W

Страница 1: ...f 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2017 SInstruments Inc S by SR Instruments Inc Ope...

Страница 2: ...CHECKLIST MODEL SRV712W 3 ASSEMBLY 4 REPLACEMENT PARTS AND ACCESSORIES 5 SYSTEM DESCRIPTION AND INTENDED USE 5 MAINTENANCE AND CLEANING 5 STORAGE AND TRANSPORTATION 6 SPECIFICATIONS 7 BUTTON FUNCTIONS...

Страница 3: ...Y 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2017 SInstruments Inc PACKING CHECKLIST Model SRV712W Marine Platform Scale DESCRIPTION QUANTITY PL...

Страница 4: ...rine Scale Platform 1 and check parts against the PACKING CHECKLIST If there are any missing or damaged parts call the Service Hotline at 1 800 654 6360 STEP 2 Place the Platform upside down on the fl...

Страница 5: ...accurate representation of the subject s weight The SRV712W Marine Platform Scale derives its power from six 6 each AA batteries With the push of a button weight is displayed with a displayed resolut...

Страница 6: ...asive cleaners The waterproof seal can also be damaged We recommend mild soap and water for general cleaning and disinfecting STORAGE and TRANSPORTATION STORAGE If storing this equipment for periods l...

Страница 7: ...dle BLUETOOTH ON TIME 5 minutes from last button press AVERAGING Automatic digital filter POWER SUPPLY Six 6 AA batteries CALIBRATION Calibration is traceable to NIST standards OPERATING CONDITIONS No...

Страница 8: ...lay box on the underside of the scale Select the mode for weighing with this Toggle Button SR SCALES APP Use the free SR Scales app Google Play store to connect with the scale from any Android device...

Страница 9: ...atically shut down after five minutes The zero TARE will be stored in memory Assist the animal on the scale platform its weight will appear on the display Note The display will automatically shut down...

Страница 10: ...four strain gauges mounted in a full Wheatstone bridge configuration These bridges convert the physical movement of the force cell due to the applied mass on the system into minute changes in electric...

Страница 11: ...selected weight from Step 1 Use the DISPLAY button to select digit positions and use button S4 to change the value STEP 7 When finished the display will read SAVE Press the DISPLAY button to save or p...

Страница 12: ...urely tighten screws NOTE TIGHTEN SCREWS ONLY UNTIL RUBBER SEAL STARTS TO COMPRESS STEP 14 Press and hold the DISPLAY button to zero the platform When the display indicates 0 0 place the calibrated we...

Страница 13: ...t and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instruments Inc be lia...

Страница 14: ...AN712W_170816 Page 14 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2017 SInstruments Inc S By...

Отзывы: