Model: SR630
i
-KG Pediatric Scale
Operating and Service Manual
Part No.: MAN630
i
-KG_200915
Page 13 of 20
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
email:
S
Instruments, Inc.
BASIC SYSTEM OPERATION (Cont’d)
WEIGHING
While the display is active, the scale will auto-weigh when the child is placed in the
cradle. The weight stable indicator “
” flashes then remains solid when stable. A
stable weight will automatically be stored into memory.
To initiate a new weigh cycle, with the child still on scale, manually press the
“
WEIGH”
button. Pressing the “
WEIGH
” button will clear the memory and take a
new reading. Once a new stable weight is indicated, the weight will be stored in
memory.
NOTE
: If there are any accessories (blanket, diaper, etc.), place them on the scale
before zeroing the system. This will ensure that the patient’s NET weight will be
displayed. It is recommended that the system be zeroed prior to each new patient.
RECALLING LAST STABLE WEIGHT
Press the recall button will display the last stable weight. A recalled weight will be
indicated by “
RCL
”; for example: “
RCL WT 12.345 Kg
”.
BATTERY REPLACEMENT
STEP 1
:
Display will read “
REPLACE BATTERIES
”.
STEP 2
: (Figure 8) Locate Battery Compartment Cover on underside of scale and unscrew the
two (2) cover screws.
STEP 3
: Remove and replace ALL six (6) “D” cell batteries. Refer to battery compartment
cover label for placement.
STEP 4
: Replace the Battery Compartment Cover and securely tighten screws.
STEP 5
: Press the “
ZERO”
button to confirm display is working.
STEP 6
: Zero the system.
Figure 8: Battery Replacement