Model SR565
i
-H Bariatric Stand-On Scale
Operating and Service Manual
Part No. MAN565
i
-H__MICROBAN_180104
Page 14 of 20
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
email:
Copyright 2018
S
Instruments, Inc.
CALIBRATION
NOTE:
Ensure that nothing is in contact with the scale system during this procedure. Remove
hands from the system when noting the displayed calibration results.
CHECKING CALIBRATION
STEP 1:
Select two (2) known calibrated weights, traceable to
NIST.
NOTE:
The first weight should be at least 500 pounds (half of the
maximum scale capacity). The second weight should be less than
half of the first weight.
DO NOT USE
barbells or un-calibrated
weights.
STEP 2:
Zero the scale by pressing and holding
ZERO / WEIGH
button.
STEP 3:
Place the first calibrated weight on the scale. Wait for
scale to stabilize; note scale reading. Remove weight.
STEP 4:
Place second calibrated weight on scale. Wait for scale to
stabilize; note scale reading. Remove weight.
STEP 5:
The scale readings for both weights should be within the
Calibration Tolerance plus or minus one digit of displayed resolution
(Figure 7).
Figure 7: Calibration Tolerance Table
Figure 8: Calibration Switch Diagram
CALIBRATION
TOLERANCE TABLE
LOW
LIMIT
APPLIED
LOAD
HIGH
LIMIT
99.9
100.0
100.1
199.8
200.0
200.2
299.7
300.0
300.3
399.6
400.0
400.4
499.5
500.0
500.5
599.4
600.0
600.6
699.3
700.0
700.7
799.2
800.0
800.8
899.1
900.0
900.9
999.0
1000.0
1001.0
IMPORTANT
CALIBRATION Qualified service personnel only should perform this procedure. The
SR565i-H load cells have no user serviceable components and should not be tampered
with for any reason. Re-calibration is generally not required, but should be verified
periodically to ensure accuracy. The recommendation for calibration check is at least
once every 12 months, or as individual maintenance policy requires.
CALIBRATION SWITCH
Continued next page