SR Instruments SR Scales SR555i Скачать руководство пользователя страница 14

Model SR555

i

 Stand-On Scale 

Operating and Service Manual 
Part No. MAN555

i

_MICROBAN_201001 

 

Page 14 of 22 

 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel:  716-693-5977     Fax:  716-693-5854    URL:  

www.srscales.com

 

email:  

[email protected]

       Copyright 2020

 

S

Instruments, Inc. 

BATTERY REPLACEMENT 

 
 
 
 
 
 
 
 
 
 
 

STEP 1:

  The display will read “

REPLACE BATTERIES

”.   

STEP 2:

  (Figure 8) Place the scale on its side and open the battery compartment cover located in 

the bottom of the scale base. 

STEP  3:

    Remove  and  replace  ALL  six  (6)  “C”  cell  batteries.    Refer  to  diagram  on  the  battery 

compartment cover for placement. 

STEP 4:

  Press the “

WEIGH

” button to confirm display is working.   

STEP 5:

  Tightly close the cover.   

STEP 6:

  Zero the system.   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 8: Bottom of Scale Base 

WARNING 

 

DO  NOT  remove  batteries  while  scale  is  on. 
Allow  scale  to  shut  down  naturally  thru 
timing  out.  This  prevents  possible  voltage 
drain  on  the  printed  circuit  board  coin 
battery. 

Содержание SR Scales SR555i

Страница 1: ...BAN_201001 Page 1 of 22 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S I...

Страница 2: ...CTION 8 STORAGE AND TRANSPORTATION 8 CLEANING AND DISINFECTING 9 SPECIFICATIONS 10 BUTTON FUNCTIONS 11 BASIC SYSTEM OPERATION 12 BATTERY REPLACEMENT 14 THEORY OF OPERATION 15 CALIBRATION 16 CONFIGURIN...

Страница 3: ...ea DISPLAY UNIT WITH CABLE FIVE 5 SET SCREWS IN PLACE 1 ea PACKAGE OF SIX 6 C CELL BATTERIES 1 ea MEDICAL GRADE AC POWER SUPPLY SR555i AC ONLY 1 ea LOCTITE 1 ea 1 8 ALLEN WRENCH 1 ea QC INSPECTION SH...

Страница 4: ...Base Assembly collar and seat firmly Re install set screws in Base Assembly collar and tighten securely NOTE If an optional printer has been ordered go to Page 6 INSTALLING OPTIONAL PRINTER before pro...

Страница 5: ...the Cable Clamp and reinstall the Cable Clamp leaving just enough slack to alleviate any tension on the Cable exiting the Platform Securely tighten the Cable Clamp screw Excess cable MUST NOT contact...

Страница 6: ...andrail Tighten all six 6 screws securely INSTALLING OPTIONAL HEIGHT BAR STEP 1 Figure 4 Remove the set screw on the back of the Display Unit 1 and in the Base Assembly 2 STEP 2 Slide the Height Bar 7...

Страница 7: ...epresentation of the patient s weight The low power microprocessor circuitry allows the SR555i to derive its power from six 6 common C cell batteries that will provide up to 10 000 weight readings bef...

Страница 8: ...dded to help reduce the growth of bacteria as part of a medical center s approach to creating a cleaner healthcare environment For more information visit https www microban com sr scales STORAGE and T...

Страница 9: ...as Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disinf...

Страница 10: ...OWER DOWN Approximately 30 seconds adjustable to 300 seconds ENTER Stores displayed reading in memory RECALL Recalls last stored stable weight height and BMI AVERAGING Automatic digital filter POWER S...

Страница 11: ...when stable Auto stores stable weight in memory SEND PRINTER EHR Press to send stored values to EHR or to printer Output values include time date weight height and BMI Also indicated on display when p...

Страница 12: ...or DOWN arrow buttons to select between PRINT INVERTED and PRINT NORMAL Press ENTER to save changes UP Press UP to adjust height up from the default to increase the scale s on time or to select a digi...

Страница 13: ...using a cane for support on the scale place the cane on the scale while zeroing the system This will ensure that the patient s NET weight will be displayed CALCULATING BMI Press to calculate BMI When...

Страница 14: ...8 Place the scale on its side and open the battery compartment cover located in the bottom of the scale base STEP 3 Remove and replace ALL six 6 C cell batteries Refer to diagram on the battery compa...

Страница 15: ...s serving to normalize the sensitivity voltage out per unit of weight applied of each bridge The offset potentiometer produces a small current which nulls the output of the amplifier for an unloaded s...

Страница 16: ...maximum scale capacity Put aside a subset of weight that is less than half of the combined weight DO NOT USE barbells or uncalibrated weights STEP 2 Zero the scale by pressing and holding ZERO WEIGH b...

Страница 17: ...Use the UP arrow button to select the digit and the DOWN arrow button to change digit Press ENTER to save changes STEP 4 Press the MENU button until HALF 500 00 Lb is displayed Change HALF to the val...

Страница 18: ...ation Mode on the display board Figure 10 STEP 6 Place the display panel into the display housing Align the display panel with the four holes in the display housing Secure display panel with the four...

Страница 19: ...Wait five seconds then re install the battery and try the WEIGH button again The display lights appear to work but do not respond to button activation Check to ensure the membrane switch label is cor...

Страница 20: ...abilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no...

Страница 21: ...and Service Manual Part No MAN555i_MICROBAN_201001 Page 21 of 22 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com...

Страница 22: ...AN555i_MICROBAN_201001 Page 22 of 22 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments...

Отзывы: