Checking the System
You may check your meter and test strips using the
TEE2 Control Solution(control A and/or B).
The TEE2 Control Solution contains a known amount
of glucose and is used to check that the meter and the
test strips are working properly.
The test strip vials have TEE2 Control Solution ranges
printed on their labels. Compare the result displayed
on the meter to the TEE2 Control Solution range
printed on the test strip vial.
Before using a new meter or a new vial of test strips,
you may conduct a control solution test following the procedure
on pages 19-20.
Notes:
•
Use only the TEE2 Control Solution (available for purchase
separately).
•
Check the expiration dates printed on the bottle. When you first
open a control solution bottle, record the discard date (date
opened plus three (3) months) in the space provided on the label.
•
Make sure your meter, test strips, and control solution are at
room temperature before testing. Control solution tests must be
done at room temperature (20-25˚C, 68-77˚F).
•
Before using the control solution, shake the bottle, discard the
first few drops and wipe the tip clean.
•
Close the control solution bottle tightly and store at a
temperature of 8-30˚C(46-86˚F).
You may do a control solution test:
•
When you want to practice the test procedure using the control
solution instead of blood,
•
When using the meter for the first time,
•
Whenever you open a new vial of test strips,
•
If the meter or test strips do not function properly,
•
If your symptoms are inconsistent with the blood glucose test
results and you feel that the meter or test strips are not working
properly,
•
If you drop or damage the meter.
Control Solution Testing
Step 1
Insert a test strip into the meter’s test strip port,
with the contact bars facing upwards.
Gently push the test strip into the port until the
meter beeps. Be careful not to bend the strip
while pushing it in.
The
symbol will be displayed on the screen.
Step 2
Shake the TEE2 Control Solution
bottle before each test.
Remove the cap and squeeze the
bottle to discard the first drop. Then
wipe the tip with a clean tissue or
cloth. After the
symbol appears
on the display, apply the solution to
the narrow edge of the test strip until the meter beeps. Make
sure the confirmation window fills completely.
LOT
Discard Date :
Manufacturer : i-SENS, Inc.
Note:
The meter may switch off, if the control solution sample is
not applied within 2 minutes of the
symbol appearing on
the screen. If the meter turns off, remove the strip, reinsert,
and start from step 1.
Содержание i-SENS TEE2
Страница 30: ...PC24 SOP OP0274 CL072 Blood Glucose Monitoring TEE2 Blood Glucose Monitors v1 November 2020 MEMO MEMO ...
Страница 31: ...PC24 SOP OP0274 CL072 Blood Glucose Monitoring TEE2 Blood Glucose Monitors v1 November 2020 MEMO MEMO ...
Страница 32: ...PC24 SOP OP0274 CL072 Blood Glucose Monitoring TEE2 Blood Glucose Monitors v1 November 2020 M E M O ...
Страница 34: ...PC24 SOP OP0274 CL072 Blood Glucose Monitoring TEE2 Blood Glucose Monitors v1 November 2020 ...