background image

RadioProcessor-G

Step 4: Specify Parameters of the Pulse Sequence

To program a pulse sequence into the RadioProcessor-G, there is one primary function that must be used:

spmri_mri_inst()

The parameters of this function are defined in Table 3.

2017-09-04

13

www.spincore.com

Table 3:

 SPMRI_MRI_INST Parameters.

Parameter

Description

double 

dac_amp

specifies the amplitude of the selected DACs on a scale of
1.0 to -1.0, where 1.0 is the highest positive amplitude 
and -1.0 is the most negative.

char

 dac_addr

selects which DACs will be updated.  Multiple DACs can
be selected at one time.

char 

 write

selects which DACs will have new amplitude data 
written.  This will not cause a new values to  be seen at 
the output.  Multiple DACs can be selected.

char

 update

selects which DACs will have updated outputs.  The 
outputs will reflect the amplitude data that was last 
written.  Multiple DACs can be selected.

char

 clear

clears the write and update registers of the selected 
DACs, causing them to cease outputting data.  Multiple 
DACs can be selected.

int

 freq

selects the frequency register that will be used for the 
carrier signal.

int

 tx_phase

selects the phase register that will be used for the carrier 
signal.

char

 tx_en

enables RF output if set to 1.  RF output will be disabled 
if set to 0.

char

 phase_reset

resets the phase to the of the carrier signal to the initial 
value.

char

 trigger_scan

enables data acquisition when set to 1.

char 

envelope_freq

selects the frequency register that will be used for the 
envelope signal.  If register 7 is used, envelope shaping 
will be disabled and the phase will be reset for the 
envelope frequency registers.

int

 amp

selects the amplitude register that will be used for the 
carrier signal.

char

 cyclops_phase

not supported on the RadioProcessor-G.  Set to 0.

UINT64 

flags

specifies the state of the digital outputs, whether high or 
low.

DWORD

 

data

the data field for the PulseBlaster opcode.

char

 op

the opcode for the PulseBlaster instruction.  See Table 3 
for more information.

double 

delay_ns

the duration of the PulseBlaster instruction.

The available instructions for the PulseBlaster core are described in detail in the RadioProcessor manual.

Содержание RadioProcessor-G

Страница 1: ...RadioProcessor G Owner s Manual SpinCore Technologies Inc http www spincore com...

Страница 2: ...ogies Inc reserves the right to make changes to the product s or information herein without notice RadioProcessor G PulseBlaster SpinCore and the SpinCore Technologies Inc logos are trademarks of Spin...

Страница 3: ...1 Step 2 Define RF Output Parameters 11 Step 3 Data Acquisition Parameters 12 Step 4 Specify Parameters of the Pulse Sequence 13 Step 5 Trigger the Pulse Program 14 Step 6 Retrieve Acquired Data 14 Us...

Страница 4: ...RadioProcessor G Hardware Trigger Reset Header JP601 20 Digital Output Header JP600 21 Related Products and Accessories 22 Contact Information 22 Document Information 22 2017 09 04 4 www spincore com...

Страница 5: ...essor manual for more details The system integrates SpinCore s high performance PulseBlaster timing engine for agile control of internal system components as well as TTL pulse pattern generation for c...

Страница 6: ...xcitation cores The gradient core controls three analog outputs that can be used to generate pulses with customizable durations and varying amplitudes of both positive and negative voltages A standard...

Страница 7: ...is shown below in Figure 2 2017 09 04 7 www spincore com Figure 2 Typical application of the RadioProcessor Model G By adding an RF power amplifier a gradient amplifier and a small signal pre amplifie...

Страница 8: ...it our website for more details Using the setup as described above with an 11 8 MHz permanent magnet a sample image of a 5 mm test tube filled with household cooking oil and two glass capillary tubes...

Страница 9: ...pling Frequency 75 1 MHz RF Analog Output D A Sampling Rate 300 MHz D A Sampling Precision 14 bits Output Voltage Range peak peak 1 2 2 V Phase resolution 0 09 deg Frequency resolution 0 28 Hz Gradien...

Страница 10: ...ent in this manual Testing the RadioProcessor G The simplest way to test whether the RadioProcessor G has been installed properly and can be controlled as intended is to run a simple test program Exam...

Страница 11: ...timing of the experiment i e pulse times delays etc 5 Trigger the pulse program The experiment will then proceed autonomously 6 Retrieve the captured data from the board at any time without interrupti...

Страница 12: ...nd in the RadioProcessor manual Once the RadioProcessor G has been instructed to acquire data there are two parameters that will influence the length for which data will be recorded these are the Spec...

Страница 13: ...e outputting data Multiple DACs can be selected int freq selects the frequency register that will be used for the carrier signal int tx_phase selects the phase register that will be used for the carri...

Страница 14: ...the SpinCore website includes a document that gives an in depth description of all the functions available in the API If you are unsure of how a particular function in an example program works please...

Страница 15: ...acquired per acquisition period nScans Number of times the entire sequence is performed nPhases Number of different phase levels to run the scan for spectrometerFrequency_MHz Frequency of the RF exci...

Страница 16: ...h all three gradients are applied Phase Gradient Time A period of time is waited to account for rise time of the gradient amplifiers Gradient Echo Delay The readout gradient coils then produce a gradi...

Страница 17: ...d gradient echo FID_Readout FID_Readout bat has delays and acquisition parameters set by default to aid in inspecting a single gradient echo This is best used to make sure that your acquisition parame...

Страница 18: ...outputs which can be controlled through the pulse program The analog output connector BNC3 is not equipped with an interpolating filter This allows for maximum flexibility in output frequency but it...

Страница 19: ...adout Gradient Hardware Trigger Reset Header JP601 The shrouded IDC header JP601 is the Hardware Trigger Reset connector This is an input connector for hardware triggering HW_Trigger and resetting HW_...

Страница 20: ...rigger is more precise since there are no software latencies involved HW_Reset pin 1 When this input is set to logical 0 for example by shorting it with pin 2 the pulse program is reset 2017 09 04 20...

Страница 21: ...nnector for header JP600 can be obtained on Digi Key Part Number 1 1658526 1 ND 2017 09 04 21 www spincore com Figure 9 Unshrouded IDC Output Header J600 Table 5 Unshrouded IDC Output Header J600 sign...

Страница 22: ...nent Magnets shtml 4 Consider complete MRI system in a small single bay enclosure Please inquire with SpinCore Technologies through our contact form which is available at http spincore com contact sht...

Отзывы: