  MECHANICAL INTERCHANGEABLE ANVIL MICROMETER 

 

                                                                                                                                           swissprec.com 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Adjustment if deviation in excess of 0.002

/ 0.05 mm:   

Use the supplied wrench to loosen the pressure ratchet (8). Lightly tap the end of the 
pressure ratchet to loosen the thimble (7) from the spindle taper. Rotate the thimble (7) 
until the zero is in alignment with the sleeve (6) base line. Carefully tighten the pressure 
ratchet (8) ensuring that the zero position is correct. 

 

 

 

 

 

Measuring Procedure:   

To measure the work piece, rotate the thimble (7) until both anvils (2) make contact with 
the feature being measured. Rotate the pressure ratchet (8) or friction thimble (not shown) 
between 1 to 1.5 revolutions to ensure that the measuring anvils (2) are correctly seated. If 
the measurement cannot be read in this position lock the spindle lock (5) and carefully 
remove the micrometer from the work piece and observe the reading. 

Parallax Error:

 

Parallax error refers to a deviation in the reading caused by not viewing the micrometer 
at right angles to the base line on the sleeve (6) 

 

 

 

Calibration: 

It is recommended that this precision measuring tool be calibrated at a periodic interval 

as follows: 

 

Annually 

 

At an interval that ensures that the accuracy is not compromised between 

calibration intervals 

 

As outlined in your company’s quality manual 

 

Changing Anvils and ensuring Zero Setting: 

Remove the anvil retainer  nut (1) from the frame (3). Insert the desired 
interchangeable anvil (2) into the frame (3). Replace the anvil retainer nut (1) 
loosely onto the frame (1). Align the pins of the anvil torque device (11) to 
the holes in the retainer nut (1). Tighten the anvil retainer nut (1) with the 
anvil torque device (11). Use the thimble      (7) to open the anvils to accept 
the appropriate mic standard (9). Place the guide rings (12) over the mic 
standard (9) leaving enough room for the  assembly to fit between the 
micrometer anvils. (2)&(4). Slide the guide rings (12) in the outer direction so 
that they are each centered where the micrometer anvils (2)&(4) make 
contact with the mic standard (9). Use the pressure ratchet to close the 
micrometer spindle (4) until it makes contact with the mic standard (9). Turn 
the pressure ratchet 1 to one 1.5 turn to ensure that the correct 
measurement can be taken. Observe the zero line on the thimble  (7) to 
ensure that it is on the base line of the sleeve (6). If necessary make 
adjustment as per special instruction 2.   

 

 

 

 

Zero Setting:   

Clean both anvils with a clean cloth to ensure both anvils are free of 
contamination. Rotate the spindle (4) by the pressure ratchet (8) until both 
anvils are both touching or making contact with the micrometer standard 
for micrometers larger than 0 to 1 inch range. Ensure that the zero on the 
thimble (7) is in alignment with the base line on the sleeve (6). 

 

Adjustment of deviation less than 0.002

/ 0.05 mm:   

Use the wrench supplied to rotate the sleeve (6) until the zero on the 
thimble (7) aligns with the base line on the sleeve. 