SPI 10-702-9 Скачать руководство пользователя страница 2

  THREE POINT INTERNAL MICROMETER 

 

                                                                                                                                           swissprec.com 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Sleeve:    7.00 mm 

Thimble: 0.29 mm 

Vernier: 0.008 mm 

Reading: 7.298 mm 

• Inch

 

In this scenario the numbers on the left repress sent the readings from the 

vernier scale. Each division counts as .0001 inch based on the division that 

lines up with the line on the thimble. 

Sleeve:    0.175

ʺ

 

Thimble: 0.004

ʺ

 

Vernier: 0.0003

ʺ

 

Reading: 0.1793

ʺ

 

Interchangeable anvil replacement instructions 

• Product measuring ranges over 4 "(100 mm) is with Interchangeable anvil.   

•  Rotate the thimble until the line marked “0” comes exactly in line with the 

reference line on the sleeve. Pull out the measuring jaws; insert another set of 

measuring jaws into the grooves in the measuring  head in turn. Inspect the 

measuring jaws by smaller pull. 

• When changing the measuring jaws, number of the measuring jaws must fit the 

relevant that of the three grooves in the measuring head. 

• Prevent dust from getting inside the three grooves in the measuring head. 

• Place tool into packaging box after apply antirust oil on the measuring jaws if the 

micrometer not used for a long of time. 

Precaution 

• Clean all measuring surfaces and body prior to use. 

• Zero set the tool using the following steps before use, and after installing the 

extension rod. 

• Check the zero point by measuring setting ring. If the reading doesn’t match the 

setting ring: 

• Unlock the adjusting screw; rotate the fastness sleeve till getting correct reading. 

• Lock the adjusting screw. 

Отзывы: