H.  Measuring Frequency 

 

1. Connect the black test lead to the "

COM" terminal and the red test lead to the "

V

Ω

Hz    

" terminal. 

 

2. Set the rotary switch in "

Hz" position. Then press the "Hz%" button until "Hz" appears on the display.

 

3. Connect the test leads across the source or circuit to be tested. 

 

4. Observe the reading on the display.

   NOTE:  

 

 The input voltage should be between 1V rms and 10V rms. If the voltage is more than 10V rms,  

the error of reading may be out of the specified accuracy range. 

I.  Measuring Duty Cycle 

 

1. Connect the black test lead to the "

COM" terminal and the red test lead to the "

V

Ω

Hz    

" terminal. 

 

2. Set the rotary switch in "

Hz" position. Then press the "Hz%" button until "%" appears on the display.

 

3. Connect the test leads across the circuit to be tested.

 

4. Observe the reading on the display.

J.  Continuity Test  

 

1. Connect the black test lead to the "

COM" terminal and the red test lead to the "

V

Ω

Hz    

" terminal.

 

2. Set the rotary switch in "         " position. Then press the 

RANGE button until the symbol "    "appears on the display.

 

3. Connect the test leads across the circuit to be tested.

 

4. If the resistance is less than about 50

Ω

, the built-in buzzer will sound and the Light Indicator will light. 

   NOTE:  

 

Before testing, disconnect all power to the circuit to be tested and discharged all capacitors thoroughly. 

K. Diode Test  

 

1.   Connect the black test lead to the "

COM" terminal and the red test lead to the "

V

Ω

Hz    

" terminal.  

(Note: The polarity of the red lead is positive "

+") .. 

 

2. Set the rotary switch in "         " position. Then press the 

RANGE button until the symbol "    " appears on the display.

 

3. Connect red test lead to the anode of the diode to be tested an 

 

4.  Read the approximate forward voltage drop of the diode on the display. If the connections are reversed,  

"

OL" will be shown on the display.

L. Auto Power Off  

 

1. 

 

If you have not operated the meter for about 15 minutes, the meter will sound 5 short beeps and a  

long beep and tum off automatically. 

 

2. 

 

If you want to turn it on again, set the rotary switch to the "

OFF" position first and then set it to a desired position. 

M.  Backlight Button   

 

1.  Press and hold down this "    " button for more than 2 secs to turn on the backlight.  

To turn off the backlight, press this button.

N.  RANGE Button   

 

1.  With the rotary switch in voltage or resistance measurement position, this "RANGE"  button can be used to  

switch between autorange mode and manual range mode as well as to select desired manual range.

 

2.  With the rotary switch in the "          " position, you can press this "

RANGE" button to switch between diode  

and continuity test functions.

O.   High Voltage Indicator   
 

1. When the hold button is press this indicator should light briefly.

 

2. If the indicator fails to light when the hold button is pressed discontinue use of the meter.

 

3.  When the AC voltage being measured is more than 750V or the DC voltage being measured is more than  

1000V, this indicator will light as a warning. 

 

4. During a continuity test this indicator will light if the resistance being measured is less than about 50

Ω

.

/

/

/

Содержание DSA200AOC

Страница 1: ...batteries are low and must be replaced 8 AC Units of Measure mV V Unit of voltage mV Millivolt V Volt 1V 1000mV mA A Unit of current µA Microamp mA Milliamp A Amper 1 A 103 mA 106 µA Ω kΩ MΩ Unit of resistance 0 Ohm kO Kilohm MO Megohm 1 MΩ 103 kΩ 106 Ω nF µF Unit of capacitance nF Nanofarad µF Microfarad 1F 106 µF 109 nF 1012 pF Hz kHz MHz Unit of frequency Hz Hertz kHz Kilohertz MHz Megahertz 1 ...

Страница 2: ...e circuit during a test If equipment under test has exposed conductive parts extra precaution should be taken to minimize the possibility of shorting Never use the instrument if its surface or your hand is wet Do not exceed the maximum allowable input of any measuring range Never open the battery cover during a measurement The instrument is to be used only in its intended applications or condition...

Страница 3: ...oming wire Interior wiring CAT III Transformer CAT II CAT I CAT IV Socket III GENERAL SPECIFICATION Display 3 3 4 digit LCD with a max reading of 3999 Overrange indication OL shown on the display Sampling rate About 2 3 times sec Negative Polarity Indication Negative sign shown on the display automatically Jaw Opening Capability 13mm Max Measurable Conductor Ø13mm Low Battery Indication shown on d...

Страница 4: ...the highest manual range first and then reduce it range by range until satisfactory resolution is obtained 3 Connect the test leads across the source or circuit to be tested 4 Observe the reading on the display E Measuring AC Current 1 Set the rotary switch in A position 2 Place the conductor wire in the jaws Make sure the wire is centered between the marks on the meter as shown in figure 3 Only p...

Страница 5: ... VΩHz terminal Note The polarity of the red lead is positive 2 Set the rotary switch in position Then press the RANGE button until the symbol appears on the display 3 Connect red test lead to the anode of the diode to be tested an 4 Read the approximate forward voltage drop of the diode on the display If the connections are reversed OL will be shown on the display L Auto Power Off 1 If you have no...

Страница 6: ...nds TIP If you want to measure a capacitor whose capacitance is less than 40nF you can measure another capacitor whose capacitance is about 1 nF and get a reading then connect the two capacitors in parallel and measure them for the total capacitance reading subtract the first reading from the total capacitance reading the result is the capacitance value of the capacitor which you want to measure T...

Страница 7: ... Soak a new swab with alcohol 4 Work the swab around in each terminal B Battery Replacement To avoid false readings which could lead to possible electric shock or personal injury replace the batteries as soon as the low battery indicator appears Before opening the case or battery cover disconnect the test leads and tum off the meter To replace the batteries remove the screws on the battery cover a...

Страница 8: ...ts original purchasers in connection with the sales of Sperry products provided that such products shall not have been altered by the distributor The distributor shall be fully responsible for any warranties the distributor makes to its purchasers which are broader or more extensive than Sperry s limited warranty Lifetime Warranty Warranty Limitation The forgoing warranties are exclusive and are i...

Отзывы: