LCD ERROR MESSAGES
The thermometer incorporates visual diagnostic messages as follows:
,
I
J
I
,LO
'Hi' or 'Lo' is displayed when the temperature being
measured is outside of the range of the instrument, 'Hi'
when higher than +250
°
C (482
°
F) and 'Lo' when lower
than -55
°
C (-67
°
F).
c_
::J
d,:
I
_,
J
'Er2' is displayed when the thermometer is exposed to
rapid changes in the ambient temperature. 'Er3' is
displayed when the ambient temperature of the
thermometer EXCEEDS 0
°
C (32
°
F) OR +50
°
C (122
°
F).
In both cases you should allow plenty of time (minimum
30 minutes) for the thermometer to stabilize to the
working/room temperature.
,f,-
BATTERIES
For all other error messages it is necessary to reset the
thermometer. To reset it, turn the instrument off, remove
the battery and wait for a minimum of one minute, reinsert
J the battery and turn on. If the error message remains
please contact the Service Department for further
assistance.
The thermometer has two separate batteries. The battery closer to the side of the laser
beam output is for laser operation only. The other battery is designated for temperature
measurements. The thermometer can still measure temperature properly, even without
the laser battery.
The thermometer incorporates visual low battery indication (for the lower battery) as
follows:
'Battery OK':
measurements
are possible
'Battery Low': battery needs to
be replaced, measurements
are possible
BATTERY REPLACEMENT
/i I I'
'Battery Exhausted':
measurements are
not possible
When the 'Low Battery' icon indicates the battery is low, the battery should be replaced
immediately with a CR2032 lithium cell. The battery is located under the twist cover at
the rear of the thermometer. Please note: It is important to turn the instrument off
before replacing the battery otherwise the thermometer may malfunction.
&
Dispose of used battery properly and keep away from children.
STORAGE & CLEANING
The sensor lens is the most delicate part of the thermometer. The lens should
be kept clean at all times, care should be taken when cleaning the lens using
only a soft cloth or cotton swab with water or medical alcohol, allowing the
lens to fully dry before using it, do not submerge any part of the thermometer.
The thermometer should be stored at room temperature between -20 to
+65
°
C (-4 to +149
°
F).
SPECIFICATION
Measurement Range
-67-482
°
F/ -55 to +250
°
C
Operating Range
32~122
°
F/ 0~50
°
C
Accuracy (Tobj=15-35
°
C,Tamb=25
°
C)
+/-1.0
°
C (1.8
°
F)
Accuracy
±2% of reading or 2
°
C (4
°
F) whichever is
(Tobj=-33-250 °C, Tamb=23±3 °C)
greater
Resolution (-9.9~199.9
°
C)
0.1
°
c
I
0.1
°
F (switchable)
Response Time (90%)
0.7 second
Distance:Spot
6: 1 optics ratio
Emissivity Range
Fixed at 0.95
Battery Life
Typ. 40hr, min 30hr
(auto power off after 15 seconds)
Battery
CR2032 *2pcs
Dimensions
0.9x2.0x4.0 inch /22.5 x 50 x 103mm
Weight
2.29 oz/ 65 grams including batteries
EMC/RFI
Readings may be affected if the unit is operated within radio frequency
electromagnetic field strength of approximately 3 volts per meter, but the performance
of the instrument will not be permanently affected.
1 Year Warranty solely to repair or replacement; no warranty of merchantability or
fitness for a particular purpose. Product is warrantied to be free of defects in materials
and workmanship for the normal life of the product. In no event shall Sperry
Instruments be liable for incidental or consequential damage.