Sperry instrument IRT100 Скачать руководство пользователя страница 2

LCD ERROR MESSAGES 

The thermometer incorporates visual diagnostic messages as follows: 

,LO 

'Hi'  or  'Lo'  is  displayed  when  the  temperature  being 

measured is outside of the range of the instrument, 'Hi' 

when higher than  +250

°

C  (482

°

F) and  'Lo' when lower 

than -55

°

C (-67

°

F). 

c_ 

::J 

d,: 

_, 

'Er2' is displayed  when  the thermometer is exposed to 

rapid  changes  in  the  ambient  temperature.  'Er3'  is 

displayed  when  the  ambient  temperature  of  the 

thermometer EXCEEDS 0

°

C (32

°

F) OR +50

°

C (122

°

F). 

In both cases you should allow plenty of time (minimum 

30  minutes)  for  the  thermometer  to  stabilize  to  the 

working/room temperature. 

,f,-

BATTERIES 

For all other error messages it is necessary to reset the 

thermometer. To reset it, turn the instrument off, remove 

the battery and wait for a minimum of one minute, reinsert 

J  the  battery  and  turn  on.  If  the  error  message  remains 

please  contact  the  Service  Department  for  further 

assistance. 

The thermometer has two separate batteries. The battery closer to the side of the laser 

beam output is for laser operation only. The other battery is designated for temperature 

measurements. The thermometer can still measure temperature properly, even without 

the laser battery. 
The thermometer incorporates visual low battery indication (for the lower battery) as 

follows: 

'Battery OK': 

measurements 

are possible 

'Battery Low': battery needs to 

be replaced, measurements 

are possible 

BATTERY REPLACEMENT 

/i  I I' 

'Battery Exhausted': 

measurements are 

not possible 

When the 'Low Battery' icon indicates the battery is low, the battery should be replaced 

immediately with a CR2032 lithium cell. The battery is located under the twist cover at 

the rear of the  thermometer.  Please note:  It is important  to turn the instrument  off 

before replacing the battery otherwise the thermometer may malfunction. 

Dispose of used battery properly and keep away from children.

STORAGE & CLEANING 

The sensor lens is the most delicate part of the thermometer. The lens should 

be kept clean at all times, care should be taken when cleaning the lens using 

only a soft cloth or cotton swab with water or medical alcohol,  allowing the 

lens to fully dry before using it, do not submerge any part of the thermometer. 

The  thermometer  should  be  stored  at  room  temperature  between  -20  to 

+65

°

C (-4 to +149

°

F).

SPECIFICATION 

Measurement Range 

-67-482

°

F/ -55 to +250

°

C

Operating Range 

32~122

°

F/ 0~50

°

C

Accuracy (Tobj=15-35

°

C,Tamb=25

°

C) 

+/-1.0

°

C (1.8

°

F) 

Accuracy 

±2% of reading or 2

°

C (4

°

F) whichever is 

(Tobj=-33-250 °C, Tamb=23±3 °C) 

greater 

Resolution (-9.9~199.9

°

C) 

0.1

°

0.1

°

F (switchable) 

Response Time (90%) 

0.7 second 

Distance:Spot 

6: 1 optics ratio 

Emissivity Range 

Fixed at 0.95 

Battery Life 

Typ. 40hr, min 30hr 

(auto power off after 15 seconds) 

Battery 

CR2032 *2pcs 

Dimensions 

0.9x2.0x4.0 inch /22.5 x 50 x 103mm 

Weight 

2.29 oz/ 65 grams including batteries 

EMC/RFI 

Readings  may  be  affected  if  the  unit  is  operated  within  radio  frequency 

electromagnetic field strength of approximately 3 volts per meter, but the performance 

of the instrument will not be permanently affected. 

1  Year Warranty solely to repair or replacement; no warranty of merchantability or 

fitness for a particular purpose. Product is warrantied to be free of defects in materials 

and  workmanship  for  the  normal  life  of  the  product.  In  no  event  shall  Sperry 

Instruments be liable for incidental or consequential damage. 

SPERRY 

INSTiUJMENTS 

The Professional's Choice® 

ACE! 