26
Apply a known concentration calibration
gas to the sensor. Toxic and LEL readings
are considered accurate if the readings
are between 90% and 120% of the
expected value as given on the gas
cylinder. If readings are accurate, then
the instrument requires no further
adjustment. Toxic and LEL readings that
do not fall between 90% and 120% of the
value on the cylinder are considered
inaccurate and indicate that the sensor
must be calibrated before further use as
discussed in sections 4.3 and 4.4.
Customers are
strongly urged to use only Sperian
calibration materials when calibrating
the NXP. Use of non-standard
calibration gas and/or calibration kit
components when calibrating the NXP
can lead to inaccurate and potentially
dangerous readings and may void the
standard Sperian warranty.
Sperian Instrumentation offers calibration
kits and long-lasting cylinders of test gas
specifically developed for easy NXP
calibration.
Press MODE to end the test and resume
normal gas reading mode.
↓
The detector will then return to the current
gas reading screen. The warning icon will
be lit to show that the alarms are off until
the readings fall below the alarm levels.
The instrument will automatically exit Test
Mode if the MODE button is not pressed
within 5 minutes.
4.6 Calibration
Failures
4.6.1 Fresh Air Calibration Failure
If the NXP is unable to complete the fresh
air calibration, the instrument will show
“no CAL” and then the countdown for Test
Mode before returning to the current gas
reading screen.
↕
4.6.2 Span Calibration Failure
If the NXP is unable to complete the span
calibration, the instrument will show “no
CAL” with the calibration cylinder icon
before returning to the current gas
reading screen.
↕