CALIBRATION
Calibration should be performed whenever the probe has been
out of use for 48 hours or longer. To prepare for calibration:
Calibrate With Temperature Sensor/ATC pH Probe
1. Press
CAL
for 2 seconds to enter Calibration Mode.
2. The meter will automatically recognize the calibration
buffer. The value and “Finished” will display on
the LCD when the procedure is complete.
3. Press
EXIT
to return to Normal Mode.
4. Repeat Steps 1-2 to calibrate other points.
Note...
All other keys will be inoperable while using the calibration function.
Calibrate Without Temperature Sensor/MTC pH Probe
1. Press
CAL
for 2 seconds to enter Temperature Setup Mode.
The temperature value will flash on the LCD.
2. Press
↑
to increase the value. Hold down
↑
to continually increase the value.
Press
↓
to decrease the value. Hold down
↓
to continually decrease the value.
For °C, the range is 0 to 60.0°C; for °F,
the range is 32 to 140.0°F.
Press
SETUP
and the meter will automatically recognize the
calibration buffer. The value and “Finished” will display on the
LCD when the procedure is complete.
Repeat Step 2 to calibrate other points.
3. Press
EXIT
to return to Normal Mode.