6.  Place the 0NTU bottle into the test chamber, lining the 

vertical white line of the bottle with the white dot on the 

meter. Close the lid and press the 

TEST/CAL

 button once. 

The meter will flash the word CAL on the screen for up to 

ten seconds.

Note… 

If the word Test appears on the screen, you did not perform 

step 6 fast enough and will need to repeat the process.
7.  If the zero point calibration was successful, the meter 

will stop blinking and display the number 100, prompting 

for the 100NTU standard solution. Again, you will have 

20 seconds to perform this step or the meter will exit 

calibration mode. 

8.  Place the 100NTU bottle into the test chamber, lining 

the vertical white line of the bottle with the white dot on 

the meter. Close the lid and press the 

TEST/CAL

 button 

once. The meter will flash CAL on the screen for up to 

ten seconds.

Note… 

If TEST appears on the screen, you did not perform the 

100NTU point fast enough and will need to repeat the process.
9.  If the calibration was successful, the screen will stop 

blinking and display 0.0, indicating that it is ready to 

test samples.

Содержание 860040

Страница 1: ...Turbidity Meter 860040 Instruction Manual...

Страница 2: ...0040 displays the turbidity level of aqueous solutions on a large LCD screen This meter can measure numerous different types of aqueous solutions and offers a two point self calibration feature that c...

Страница 3: ...use in the field or laboratory Data Hold Function Optional DC adapter for field use Zero adjustment feature for low turbidity solutions MATERIALS SUPPLIED Turbidity meter Calibration standards solutio...

Страница 4: ...ton 7 Zero Button 8 Min Max Button 3 FRONT PANEL DESCRIPTION 3 1 Cover of Testing bottle 3 2 Container of Testing bottle 3 3 Display 3 4 Hold Button Esc Button 3 5 TEST CAL Button 3 6 Power Button 3 7...

Страница 5: ...ver 2 Remove the battery cover 3 Insert six new AAA batteries ensuring correct polarity 4 Replace the battery cover and reinstall the screws Replace the batteries when the low battery icon blinks on t...

Страница 6: ...ten both caps and shake the bottles gently to ensure they are properly mixed 2 Clean the outside of both bottles to ensure there are no fingerprints on the glass Fingerprints or dirt on the glass will...

Страница 7: ...d display the number 100 prompting for the 100NTU standard solution Again you will have 20 seconds to perform this step or the meter will exit calibration mode 8 Place the 100NTU bottle into the test...

Страница 8: ...the minimum and maximum values of a set of samples This is especially useful in the case where there are slight changes in the turbidity of many samples or there is heavy sediment in a single sample...

Страница 9: ...ity systems such as well water tap water or surface water from lakes and rivers This method is effective for any solutions whose expected value is 2 0 NTU The feature allows you to define a custom sta...

Страница 10: ...All readings will be offset by the zero value until the meter is recalibrated or zeroed with the 0NTU standard Data Hold The Minimum Maximum functions on the meter are disabled during data hold Also n...

Страница 11: ...on alcohol or benzenes on the meter Repairs or services not covered in this manual should be performed by qualified personnel only Please contact Sper Scientific to speak with a technician Periodicall...

Страница 12: ...ot cover sample vials batteries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will break the waterproof seal an...

Отзывы: