6
9. “SA” (save) is displayed as the calibration point
is saved.
10. “En” (end) (850050, 850051), or “End”
(850052) displays when the calibration ends.
11. For 2 or 3 point calibration, start with pH 7,
then repeat the above steps for pH 4 and/or
pH 10.
12. For the most accurate readings, the
temperature of the sample and cliabration
solutions should be similar.
13. Repeat the above procedures until
readings stabilize.
14. NOTE: If the buffer needs to be adjusted, press
▲ to adjust the displayed value. “SA” (save) is
displayed, followed by “En” (850050, 850051),
or “End” (850052).
MEASUREMENT PROCEDURE 850050
1. Calibrate routinely before use.
2. Remove the protective cap and clean the
electrode with de-ionized/distilled water. Do not
touch the glass bulb.
3. Press >1 sec to turn on the unit.