III.  OPERATING INSTRUCTIONS
A. PH CALIBRATION PROCEDURES 

The meter is factory calibrated to simulate the ideal pH electrode  

(0 mV at pH 7.00 and 177.4 mV at pH4, based on an ambient 

temperature of 77°F/25°C ).  Follow the two-point calibration procedures 

to ensure the highest degree of accuracy.

Two Point Calibration 

1. Connect the pH electrode to the 

BNC SOCKET

.

2. Press the 

ON

 button and place the pH electrode into the pH7 solution.

3.  Using a small screw driver, turn the 

CAL. PH7

 adjustment until the 

display indicates 7.00.

4. Rinse the electrode in distilled water and shake dry.
5. Next, place the electrode into the pH4 solution.
6.   Turn  the 

SLOPE PH4 PH10

 adjustment until the display  

indicates 4.00.

7. Rinse the electrode again.

Single Point Calibration

1. Connect the pH electrode to the

 BNC SOCKET

.

2. Press the 

ON

 button and place the pH electrode into the pH7 solution. 

3.  Using a small screw driver, turn the 

CAL. PH7

 adjustment until the 

display indicates 7.00.

4. Rinse the electrode in distilled water. 

B. PH MEASUREMENT PROCEDURES

1. Connect the pH electrode and press the 

ON

 button.

2. Calibrate the meter (see above instructions).
3. Place the pH electrode into the solution to be measured.  
4. The pH level of the solution is displayed.
5.  To freeze the displayed value, press the 

HOLD

 button.   

To cancel this function, press the 

HOLD

 button again.

6. Rinse the electrode in distilled water after each use.
7. Press the 

OFF

 button.