To return to recording mode, press the 

HOLD 

(4) button. "REC" is  

displayed without "Max" or "Min". 

To erase the recorded Max/Min values and exit the recording mode, 
press the 

REC MAX/MIN 

(5) button for at least 2 seconds. 

 
3-E.  

Notes and Precautions 

 

Avoid range overload.   

When using PEAK HOLD (3-C) or REC MAX/MIN (3-D) your  
measurements must be within the selected range, if not, the out of 
range indicator "- - - -" will be displayed. 

Cover the 

PHOTO SENSOR 

(16) with the sensor cap when the meter 

is not in use. 

Fluctuations in the reading are often due to shadows or a weak bat-
tery.  Also, ambient temperatures and drafts may affect the reading. 

Do not store in areas of high temperature and/or humidity.   

Remove the battery for long-term storage. 

Do not adjust the calibration trimmer beneath the 

RS232 Cover 

(6).  

 

4.  ZERO ADJUSTMENT AND CALIBRATION 

 

With the sensor cover securely in place and the meter turned on, use 
the 

RANGE 

(9) button to  select the 40.00 Lux range (digits appear).  

Press the 

ZERO 

(11) button and the "

0.00 

LUX

" should be displayed. 

To maintain accuracy, annual laboratory calibration is recommended 
and available from Sper Scientific. 

 

5.  BATTERY REPLACEMENT 

 

Replace the battery when the low battery icon is displayed. 

In-spec measurements may be made for several hours after the low 
battery indicator appears.   

Remove the 

BATTERY COVER 

(14) screw and slide the cover off.   

Replace the battery with a fresh 9V battery (alkaline or heavy duty). 

 

6.  RS232 PC SERIAL INTERFACE 

 

Use a small screwdriver to gently lift the 

RS232 COVER 

(6) up to expose 

the 

RS232 OUTPUT 

(8) 3.5 mm terminal.  The signal output is a 16-digit 

data stream that can be adapted to user-defined applications.

  

A RS232 

lead with the following connection is required to link the instrument with the 
PC serial interface.   

 

Meter (3.5 mm jack plug)             PC (9W ‘D” Connector) 
Center Pin .................................. Pin 4      Pin 2      2.2 K 
Ground/shield ............................ Pin 2      Pin 5      resistor

 

[email protected]

www.ponpe.com

Содержание 840022

Страница 1: ...Advanced Light Meter 840022 Instruction Manual SPER SCIENTIFIC LTD sale ponpe com www ponpe com ...

Страница 2: ...ccurate response Results are easily read on the extra large LCD with bar graph Measures Foot Candles or Lux and offers high resolution 5 ranges selectable light source types Tungsten Fluorescent Sodium or Mercury max min peak hold RS232 output zero adjustment a tripod back automatic power off low battery and out of range indicators The sensor is cosine and color corrected and hermetically sealed t...

Страница 3: ...ZERO SOURCE 15 16 1 Sensor Input 2 Display 3 Power Button 4 Hold Button 5 REC MAX MIN Button 6 RS232 Cover lifts 7 Lux FC Button 8 RS 232 Output see 6 9 Range Button 10 Source Button 11 Zero Button 12 Tripod Screw 13 Stand 14 Battery Cover 15 Sensor Plug 16 Photo Sensor sale ponpe com www ponpe com ...

Страница 4: ... life After approximately 10 minutes without activity no buttons pushed the meter will automatically shut off To disable this feature press the REC MAX MIN 5 button Press the POWER 3 button to manually turn the meter off 3 B Hold During measurement press HOLD 4 to freeze the measured value HOLD and the value are displayed Press the HOLD 4 button again to exit 3 C Peak Hold During measurement press...

Страница 5: ...ecurely in place and the meter turned on use the RANGE 9 button to select the 40 00 Lux range digits appear Press the ZERO 11 button and the 0 00 LUX should be displayed To maintain accuracy annual laboratory calibration is recommended and available from Sper Scientific 5 BATTERY REPLACEMENT Replace the battery when the low battery icon is displayed In spec measurements may be made for several hou...

Страница 6: ... USB Adaptor Cable 840090 Water Resistant Instrument Pouch 840092 Bench Top Tripod 840093 Field Tripod 850080 Software The 16 digits data stream will be displayed in the following format D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0 Each digit indicates the following status D0 End Word D1 D8 Display reading D1 LSD D8 MSD E g If the display reading is 1234 then D8 to D1 is 00001234 D9 Decim...

Страница 7: ...F S 400 0 36 6 400 0 0 1 4 000 360 4 000 1 40 000 3 600 40 000 10 400 000 10 000 400 000 100 100 000 3 rdg 0 5 F S 100 000 for reference only Foot Candle 4 000 0 3 720 0 001 3 rdg 0 5 F S 40 00 3 35 37 20 0 01 400 0 33 5 372 0 0 1 4 000 335 3 720 1 40 000 930 37 200 10 9 300 3 rdg 0 5 F S 9 3000 for reference only Accuracy tested by a standard parallel light tungsten lamp of 2856 K sale ponpe com ...

Страница 8: ... substituted if available This warranty does not cover probes batteries or damage resulting from accident misuse or abuse of the product To obtain warranty service ship the unit postage prepaid to SPER SCIENTIFIC LTD 7720 E Redfield Rd Suite 7 Scottsdale AZ 85260 WWW SPERSCIENTIFIC COM INFO SPERSCIENTIFIC COM The defective unit must be accompanied by a description of the problem and your return ad...

Отзывы: