- 5 -
electrical calibration of the unit.
•
The LCD displays "CAL" for about 1
second. As needed, use the calibration
tool to adjust the Calibration (CAL) VR
until "94.0dB" is displayed.
•
This ends the calibration. Make sure
that the
Measurement Slide Switch
is
returned to the "MEAS" (up) position
and the
Battery Cover
is reattached.
2. Acoustic calibration using a standard
sound calibrator such as Sper Scien-
tific's 840031.
•
With the Measurement Slide Switch in
the "MEAS" (up) position, place the mi-
crophone in the calibrator.
•
As needed, use the calibration tool to
adjust the Calibration (CAL) VR until
"94.0dB" is displayed.
Measurement
1. Press the
Power / Max Hold
button to
turn the meter on.
2. Point the
Microphone
at the noise
source.
3. The LCD displays the measured sound
level. Readings are updated twice
every second.
•
"OVER" is displayed when the sound