SOUND LEVEL PEN 

 

840018 

 

INSTRUCTION 

MANUAL 

 
 
 

SPER  
SCIENTIFIC 

Содержание 840018

Страница 1: ...SOUND LEVEL PEN 840018 INSTRUCTION MANUAL SPER SCIENTIFIC ...

Страница 2: ...SI S1 4 Type 2 and IEC61672 1 Class 2 Meas ures the 32 130 dB range in the A scale simulating the human ear response Features a built in calibration adjustment fast response time mini size a resolution of 0 1 dB Max Hold and a low battery indicator Designed to meet the measurement requirements of safety engi neers health industrial safety offices and sound quality control in various environ ments ...

Страница 3: ... LR03 1 5V X 2 5 6 Power Off Press 2 Sec SOUND 840018 LEVEL PEN Max Hold 1 Windscreen 2 LCD Display 3 Power On Off Max Hold button 4 Microphone 5 Measurement Slide Switch 6 Calibration CAL VR 7 Battery Cover II Description www ponpe com ...

Страница 4: ...ong period of time Wipe the case with a dry cloth do not use abrasives or solvents Calibration Calibrate the meter before first use or when the reading appears inaccurate If you require a certificate of calibration traceable to NIST the unit may be sent to Sper Scientific for servicing fee applies There are two methods of calibration 1 Internal Electrical Calibration With the unit on lightly press...

Страница 5: ...oustic calibration using a standard sound calibrator such as Sper Scien tific s 840031 With the Measurement Slide Switch in the MEAS up position place the mi crophone in the calibrator As needed use the calibration tool to adjust the Calibration CAL VR until 94 0dB is displayed Measurement 1 Press the Power Max Hold button to turn the meter on 2 Point the Microphone at the noise source 3 The LCD d...

Страница 6: ... will be displayed Press the Power Max Hold button again to exit this function 5 To turn the unit off press and hold the Power Max Hold button for 2 seconds Battery Replacement Replace the battery when the low battery icon is displayed press the Battery Cover to slide it down and off Insert two 1 5V batteries paying attention to the polarity as indicated in the compartment Reattach the cover Optio...

Страница 7: ...Operating Temperature 32 104 F 0 40 C Operating Humidity 10 to 90 RH Storage Temperature 14 140 F 10 60 C Storage Humidity 10 to 75 RH Display Update 0 5 seconds Microphone 1 2 electret condenser microphone Dimensions 8 1 L 1 7 W 1 0 H 206 42 25 mm Weight 0 22 lbs 100g with batteries Battery Life Approx 20 hours Alkaline IV SPECIFICATIONS www ponpe com ...

Страница 8: ...ct will be substituted if available This warranty does not cover damage resulting from accident misuse or abuse of the product To obtain warranty service ship the unit postage prepaid to SPER SCIENTIFIC LTD 7720 E Redfield Suite 7 Scottsdale AZ 85260 WWW SPERSCIENTIFIC COM INFO SPERSCIENTIFIC COM Please Note The defective unit must be accompanied by a description of the problem and your return add...