Sper scientific 840006 Скачать руководство пользователя страница 3

3

PB

3.  MEASUREMENT PROCEDURES

• Insert the 

SENSOR

 plug into the 

SENSOR INPUT

 socket.

• Turn the meter on using the 

ON 

button.  

• Remove the lens cap from the

 PHOTO SENSOR.

•  Press the 

1X

 button for measurements of 0 to 1999 LUX, or select 

the 

10X

 button for measurements of 2,000 to 19,990 LUX. If you are 

uncertain about which range to use, select the highest range.  

In 10X mode, “

  

 “ appears in the right bottom LCD.

•  Range Overload is indicated when “1” appears on the left side of the 

display.  If this occurs, switch to the highest range or discontinue use.

•  Point the

 PHOTO SENSOR

 toward the source and read the results 

on the display. To convert LUX results to FC, multiply the meter 

reading by the factor of .0929

•  Push the 

HOLD

 button to freeze the reading in the display. The word 

“HOLD” is displayed. Press the 

HOLD

 button again to exit this function.

• Press the

 OFF 

button to turn the unit off.

NOTES:

•  Keep the lens cap on the PHOTO SENSOR at all times except when 

actually taking a reading.

•  Fluctuations in the reading are generally due to shadows or 

fluctuations in the line voltage. Ambient temperature, drafts and 

ventilation also affect the luminous flux output.

•  Avoid range overload and do not store in areas of high temperature  

and/or humidity. Remove the battery for long-term storage.

4. ZERO ADJUSTMENT & CALIBRATION 

•  With the sensor cover securely in place and the meter turned on, the 

LCD display should indicate “0000”.  A slight variation will not affect 

readings; however, if a precise reading is required you may use the 

ZERO SETTING

 terminal to reset the unit.  

•  To make an adjustment, cover the 

PHOTO SENSOR

 and press  

the 

1X

 button.

•  Use a small screwdriver to turn the 

ZERO SETTING

 terminal until 

“0000” is displayed. 

•  To maintain accuracy, annual laboratory calibration is recommended.

WARNING:

 

Only use the terminal marked "ZERO" for adjustment. 

The unmarked terminals are for professional laboratory calibration only 

and tampering with these terminals will result in inaccurate readings 

and require professional calibration of the meter. 

Отзывы: