OPERATING PROCEDURE

 

1.  With the 

COVER PLATE 

open, clean the 

PRISM 

with a 

soft cloth to avoid scratching the surface.

 

2. Aim the refractometer toward a light source and rotate 

the 

EYEPIECE 

to obtain the clearest focus.

 

3.  Zero Point Calibration:

 

A.  Open the 

COVER PLATE

.

 

B. Apply a few drops of the included Distilled Water 

onto the 

PRISM 

platform.

 

C.  Close the 

COVER PLATE

.

 

D. Turn the 

CALIBRATION SCREW 

until the dark and 

light boundary line coincides with the bottom line

 

of the scale.

 

4.  Carefully dry the prism platform and 

COVER PLATE

.

 

5.  Place a few drops of the test solution on the prism and 

close the 

COVER PLATE 

so the solution spreads evenly 

on the prism.

 

6. Aim the front of the refractometer towards the light 

source and focus the eyepiece on the boundary line.

 

7.  The boundary line indicates the v/v percentage of 

alcohol in the test sample at 20°C.

 

8. After use, clean the prism with distilled water and a cloth 

to remove any residue.

 

 

Note:

 The temperature of the zero point calibration 

should be at the same temperature as the test solution. 

For variations in temperature the zero point should be 

adjusted once every 30 minutes.

 

 

Note:

 For beer or wine, this unit is to be used in conjunction 

with a hydrometer before and after fermentation process. 