3.  OPERATING INSTRUCTIONS 
 
3-A.  General Measurement 

 

• 

To avoid damage, it is important to clean the 

PRISM  

WINDOW

 before and after use. 

• 

Press the 

ON/OFF 

button to turn the Digital Refractometer 

on.  The unit will automatically shut off after about 3  
minutes of inactivity. 

• 

Perform the 

Calibration Procedures 

(below). 

• 

With the unit on, fill the 

PRISM WINDOW 

with the sample 

to be measured, then close the 

COVER

• 

Press the 

READ 

button immediately to ensure accurate 

readings. Delays may cause the sample to evaporate. 

 

3-B.  Changing the Temperature Scale:  

With the unit on, 

press the 

FUNC 

button as needed until 

l.P  C 

(or 

F

 

) is  

displayed.  Press the 

SCALE 

button to view the temperature 

in Fahrenheit (or Celsius).  Press the 

READ 

button to exit.

 

 
3-C.  Automatic Temperature Compensation (ATC):  

This 

feature makes manual temperature correction unnecessary.  
However, to obtain precise readings the sample and the unit 
should be at similar ambient temperatures.  ATC appears on 
the LCD when this function is activated.  To set the ATC  
function, press the 

FUNC 

button as needed until 

ALC.Y 

(or 

n

)  

appears.  Press the 

SCALE 

button to toggle between on or 

off.  Press the 

READ 

button twice to exit. 

 
3-D.  Calibration Procedures 

 

• 

Clean the 

PRISM WINDOW 

before and after every use. 

• 

Fill the 

PRISM WINDOW 

with a standard solution such as 

distilled or deionized water.  The ideal temperature of the 
solution and the unit should be 65°~72°F (18°~22°C). 

• 

Close the 

COVER

• 

Press the 

ON/OFF 

button to turn the unit on.  

• 

Press the 

ZERO 

button for 3 seconds. 

• 

When the procedure is complete, CAL will be displayed. 

• 

To check the nD value, press the 

ZERO 

button once. 

• 

This data can be viewed at any time to show the last  
calibration value. 

• 

A flashing crosshair symbol is displayed when the  
calibration solution falls outside the accepted range. 

Содержание 300033

Страница 1: ...DIGITAL REFRACTOMETER 300033 300034 300035 300036 INSTRUCTION MANUAL ...

Страница 2: ...tory accuracy The digital readout eliminates uncertainties while the water resistant design facilitates clean ups and protects the electronics from moist environments All models display sample temperature in addition to their other parameters Refractometers operate on the principle that as the concentration or density of a solution increases its refractive index changes proportionately The larger ...

Страница 3: ...Cover FACEPLATE LCD DISPLAY 1 Prism Temperature 2 ATC activated when displayed 3 High Limit Alert 4 Low Limit Alert 5 Low Battery 6 Measurement Value 7 Calibration Error 8 Scale Indicator 2 PANEL DESCRIPTION 1 2 3 4 5 6 7 8 9 10 Brix nD 6 7 8 nD 76 4 F ATC H Limit L Limit 1 8 8 8 8 READ ZERO SCALES FUNC ENTER Y N ON OFF 1 2 3 4 5 ...

Страница 4: ... manual temperature correction unnecessary However to obtain precise readings the sample and the unit should be at similar ambient temperatures ATC appears on the LCD when this function is activated To set the ATC function press the FUNC button as needed until ALC Y or n appears Press the SCALE button to toggle between on or off Press the READ button twice to exit 3 D Calibration Procedures Clean ...

Страница 5: ...t on press the FUNC button as needed until L L Y and the flashing L Limit icon is displayed Press the READ button and the last digit flashes Use the and the buttons to set the low range limit Press the READ button activate the function Press the FUNC button as needed to return to the measurement screen Deactivating the Low Range Limit Alert Press the FUNC button as needed until L L n appears Press...

Страница 6: ...etergent and water Do not use in corrosive environments Do not submerge the unit This precision measuring instrument contains delicate optics and electronics Handle with care and do not drop To avoid electric shock make sure the battery compart ment is covered and do not disassemble the unit 3 J Erratic Readings If the readings are significantly above or below the expected results consider the fol...

Страница 7: ...T 0 280 1 1 300036 USG 1 000 1 090 0 001 0 002 Serum Protein 0 0 12 0 0 1 0 1 nD 1 330 1 4098 0 0001 0 001 Temperature All Models 0 40 C 0 1 C 1 C 32 104 F 0 1 F 2 F nD 1 3330 1 4419 0 0001 0 001 Power Supply 9V Battery apx 5000 tests battery Weight 18 oz 500 g Dimensions 7 x 4 x 2 180 x 100 x 65 mm Operating Temp 41 95 F 5 35 C Storage Temp 14 122 F 10 50 C Included Protective Case 9V Battery Dis...

Страница 8: ...lable This warranty does not cover probes batteries or damage resulting from accident misuse or abuse of the product In order to obtain warranty service simply ship the unit postage prepaid to SPER SCIENTIFIC LTD 7720 East Redfield Suite 7 Scottsdale Arizona 85260 480 948 4448 info sperscientific com www sperscientific com Please Note The defective unit must be accompanied by a description of the ...

Отзывы: