System Purging
Carefully inspect the system before initial start-up. Check that all plumbing and electrical connections are
not loose or have come undone during shipment. A filter housing wrench is included with the system.
NOTE:
Leave the power to the system off for this procedure.
1. Redirect permeate water to the drain for this procedure.
2. Fully open the concentrate valve
(Figure 1, Item 5, Counter Clockwise).
3. Fully close the concentrate recycle valve
(Figure 1, Item 6, Clockwise).
4. Fully open the throttle valve
(Figure 1, Item 18, Counter Clockwise).
5. To activate the solenoid bypass feature, place the three way push button located on the front panel of
the C-23 controller to the FLUSH position
(Figure 3).
6. Turn the feed water on the let the system purge until no visible bubbles appear from concentrate flow
meter.
7. Place the three way push button in OFF position
(Figure 3).
Membrane Installation,
Removal & Replacement