8
Meter Configuration
The digital reader for the Field Scout EC meter is
shipped in Manual, 1-Point calibration, HI range
mode. In this mode, the EC units will always be
displayed as mS/cm. If you are working with less
concentrated solutions and want to calibrate or read
at a lower concentration, the range mode should be
changed to one of the other modes. This will allow
the meter to read in units of
μ
S/cm.
Configuring the calibration mode
1. Switch off the display.
2. Press and hold the
INC
key (see
p. 6 for diagram
of calibration keys) then simultaneously press the
ON/OFF
button to switch on the display. The meter
will go into calibration selection mode.
3. The lower display shows "A.CAL" and the upper
display blinks the current choice (YES or NO). Press
either the
INC
or
DEC
key to select the NO option.
This disables the Automatic calibration mode and
configures the display for Manual calibration.
4. Press the
HOLD/ENTER
button to confirm the
selection. The display shows "CO".
5. After 1 second, the lower display will show "1.Pnt"
and the upper display blinks the current choice (YES
or NO). Press either the
INC
or
DEC
key to select
the YES option. This enables the 1-Point calibration
option.
4. Press the
HOLD/ENTER
button to confirm the
selection. The display shows "CO" and, after a few
seconds, goes to measurement mode.