Spectrum Field Scout EC 450 Скачать руководство пользователя страница 11

11

 

Tip:  If  you  have  your  GPS  unit  set  properly  and 

have checked the connection but still are not get-

ting  geo

-

referenced  data,  uncheck  the  box  requir-

ing the digital correction in Meter Settings (p. 8).

 

Initial GPS connect

 

If the meter does NOT detect an incoming GPS signal, 

it  will  flash  steadily  for  10  seconds  and  then  glow 

steadily once it gets into 

Measurement

 mode. If it does 

detect the GPS signal (satellites found etc..), it gives a 

series  of  3  rapid  flashes  before  going  to  the  steady 

glow  for 

Measurement

  mode.  If  a  GPS  signal  is  not 

found during power

-

up, the meter  will not search for it 

when  taking  readings  and  no  geo

-

referenced  data  will 

be collected during that sampling session. If the meter 

is  turned  off  and  back  on,  it  will  again  search  for  the 

GPS  signal.  Be  sure  the  meter  is  in 

Measurement 

mode (see p. 4) before taking any readings.

 

 

Taking measurements

 

If  the  meter  did  NOT  detect  the  GPS  initially,  once  in 

Measurement

  mode,  the  LED  will  go  off  and  on  very 

quickly when a measurement is taken.

 

 

If  it  did  detect  the  GPS  signal  initially,  once  in 

Meas-

urement 

mode,  it  will  look  for  the  GPS  signal  at  the 

time of every measurement. There are two possibilities:

 

 

1. If it sees the GPS signal at the time of measurement, 

the LED will go off for a noticeably longer time and then 

come back on as a steady glow. Latitude and longitude 

values will be added to the data file. The meter is again 

ready  to  take  a  reading  when  the  LED  returns  to  a 

steady glow. 

 

2.  If  it  does  NOT  see  the  GPS  signal  at  the  time  of 

measurement  (signal  lost,  batteries  die,  cable  gets 

pulled  out),  the  LED  will  flash  briefly  and  then  come 

back on as a steady glow .

 

Содержание Field Scout EC 450

Страница 1: ...EC 450 Meter Data Logger PRODUCT MANUAL Item 2220WP ...

Страница 2: ...rder call Spectrum Technologies Inc 800 248 8873 or 815 436 4440 between 7 30 am and 5 30 p m CST FAX 815 436 4460 e mail specmeters com www specmeters com Spectrum Technologies Inc 3600 Thayer Court Aurora IL 60504 General Overview 3 Data Logger Operation 4 Identifying the Correct Com Port 6 Connecting to a PC 7 Logger Software 8 GPS Connection 10 Data Files 12 Specifications 13 Service and Suppo...

Страница 3: ... data with the EC 450 meter The logger can be used with or without GPS DGPS After a set of data has been collected it is transferred to a PC using the included software The data is in ASCII text file format and can be exported into mapping software or popular spreadsheet software for analysis NOTE THIS BOOKLET CONTAINS INSTRUC TIONS ON OPERATING THE GPS COMPATIBLE DATALOGGER AND MAINTAIN ING THE E...

Страница 4: ...is attempted during this 10 second period the logger will remain in Communication mode indefinitely The meter must be turned off and then on to change to Meas urement mode see following section Measurement Mode Approximately 10 seconds after turning on the meter the PRINT button will glow steadily This indicates the logger is in Measurement mode In this mode you can calibrate the meter and record ...

Страница 5: ...dScout software The GPS compatible log ger is operated as follows When the probe is inserted in soil or water the me ter s LCD will immediately begin indicating the EC When this value stabilizes the meter locks on to that value and the LCD will display READY in the lower left portion of the screen To capture this value with the data logger press the ENTER HOLD button followed by the PRINT button T...

Страница 6: ...device such as a modem is probably connected to that port If the message No Connection is displayed this port may be the one connected to your serial cable and you can proceed to the next step 3 Place a paperclip on the end of the serial pin so that it touches both the tip of the pin and the metal area between the two black rings Again click on the Test Port Now button If the message Connection OK...

Страница 7: ...puter port The port is locat ed at the base of the meter behind a rubber cover marked OUTPUT The meter s configuration can be modified by clicking on the Meter Settings button see Meter Set tings p 8 The Com Port Meter Type Download Clear Memory and Meter Settings buttons are explained in the Logger Software section p 8 Connecting to a PC ...

Страница 8: ...llows you to configure the data logger The Meter Name will be the title on the first line of the downloaded files If the box below the logger name field is checked the logger will store GPS data only if it has been dif ferentially corrected If the differential correction is not found only the EC reading will be stored in the data file It is recom mended that this box be left unchecked An adjustmen...

Страница 9: ...o save it The folder selection field on the right allows you to browse to any folder in your system When the file has been saved the software will give you the option of immediately viewing the file The data file is stored as a comma delimited text file and may be viewed in any text editor or spreadsheet soft ware Clear Memory Data is not automatically removed from the logger memory after a downlo...

Страница 10: ... your GPS unit it should be available from the manufacturer This cable is generally used to upload information from a computer to the GPS unit These components should be connected as shown in the figure below GPS Setting Your GPS unit must be set for NMEA 0183 input output messages If the meter has trouble receiving the GPS signal check that it has the following set tings Data bits 8 Stop bits 1 B...

Страница 11: ... and back on it will again search for the GPS signal Be sure the meter is in Measurement mode see p 4 before taking any readings Taking measurements If the meter did NOT detect the GPS initially once in Measurement mode the LED will go off and on very quickly when a measurement is taken If it did detect the GPS signal initially once in Meas urement mode it will look for the GPS signal at the time ...

Страница 12: ...ferenced to the 1984 World Geodetic Survey datum The fourth line shows the column head ings for the rest of the data file Logging sessions are started and completed by turning the meter on and off The start of a logging session is in dicated by the data line Logger Started If a GPS signal was found at the start of a logger session a time stamp is included on the Logger Started line The data is sep...

Страница 13: ...a points without GPS 1 488 data points with GPS Fresh batteries provide 40 hours of logging Operating Environment Weather Resistant Software Requirements Windows 95 or higher Field Scout Software v 3 4 or higher included Specifications ...

Страница 14: ...ake a few moments to get the answer you need 2 Write down the events that led to the problem Have you changed anything in your computer recently Are you doing anything differently When Contacting Spectrum Technologies Inc please indicate that you need Technical Support Be prepared to 1 Provide details on the hardware and software configuration of your computer including manufacturer model number p...

Страница 15: ...This warranty does not cover damage due to improper installation or use lightning negligence accident or unauthorized modifications or to inci dental or consequential damages beyond the Spectrum prod uct Before returning a failed unit you must obtain a Returned Materials Authorization RMA from Spectrum Spectrum is not responsible for any package that is returned without a valid RMA number or for t...

Страница 16: ...ription Data Logging EC Meter Type Electrical Equipment for Measurement Control and Laboratory Use Directive 89 336 EEC Standards EN 61326 1997 EN 55022 EN 61000 4 2 3 The meter contains an add on data logger that has been validated to meet CE certification Douglas L Kieffer Soil Water Products Manager March 4 2009 3600 Thayer Court Aurora IL 60504 800 248 8873 or 815 436 4440 Fax 815 436 4460 E M...

Отзывы: