Operation 

Power On/Off 

Press the On/Off/Set key to turn the meter on or off.  During 
warm-up, it will perform a 30 second countdown, then will dis-
play current CO

2

, temperature, and humidity.  The meter 

turns off automatically after 20 minutes of inactivity.  To bring 
the meter to non-sleep mode, press and hold SET and HOLD 
for 2 to 3 seconds.  You should see a character (n) appear on 
the LCD display.  At this point, let go of the buttons to cancel 
the automatic power-off function. 
 

Taking a Measurement 

The meter measures continuously and updates the LCD 
every second.  With a significant environmental change, the 
CO

2

 sensor requires 30 seconds, and the RH sensor 30 min-

utes to respond.  Note: Do not hold the meter close to the 
face because exhalation affects the CO

2

 reading. 

 

Switch Temperature Measurement 

Press the DP/WBT key to cycle the temperature display be-
tween air, dew point, and wet bulb temperatures. 
 

Data Hold 

Press the HOLD button to freeze the readings.  HOLD will 
display on the LCD.  Press HOLD again to cancel the hold 
function. 
 

Backlight 

Hold the MODE button for more than 1 second to activate 
and cancel the backlight function. 
 

MIN, MAX, STEL, TWA 

Press the M

N

X

/AV button to switch between the normal, mini-

mum, maximum, and weighted average readings. 
 
In MIN and MAX modes, minimum and maximum readings 
for CO

2

, air/dew point/wet bulb temperature, RH will display.   

In STEL and TWA modes, the display shows the weighted 
average of the CO

2

 readings for the past 15 minutes (STEL) 

or 8 hours (TWA), as well as the current readings for air/dew 
point/wet bulb temperature, and RH. 

Содержание 3440

Страница 1: ...Temp RH CO2 Hand Held Meter PRODUCT MANUAL Item 3440 ...

Страница 2: ...iation from 1000hPa Range 14 to 140 F 10 to 60 C Accuracy 0 9 F 0 6 C Humidity Range 0 to 99 9 RH Accuracy 3 10 90 5 other Warm up 30 seconds Power 4 AA batteries or 9VDC 1A adaptor not incl Battery life 24 hours alkaline Temperature Thank you for purchasing the Temp RH CO2 Hand Held Meter With an NDIR non dispersive infrared CO2 sensor this unit is stable for long term monitoring Please read this...

Страница 3: ... the battery icon and Lob will appear on the LCD and the beeper will sound press any key to stop the beeps Symbols HOLD Readings are frozen TWA Time weighted average 8 hours STEL Short term exposure limit 15 min wt avg MIN MAX Minimum Maximum readings DP Dew point temperature AIR Air temperature WBT Wet bulb temperature Battery LCD Display Keypad ...

Страница 4: ...e Do not hold the meter close to the face because exhalation affects the CO2 reading Switch Temperature Measurement Press the DP WBT key to cycle the temperature display be tween air dew point and wet bulb temperatures Data Hold Press the HOLD button to freeze the readings HOLD will display on the LCD Press HOLD again to cancel the hold function Backlight Hold the MODE button for more than 1 secon...

Страница 5: ... up arrow to increase the value or DP WBT down arrow to decrease it Press the MN X AV but ton to save the value or CAL ESC to exit without saving The meter features audible alarm to warn by emitting 80dB beeps when the CO2 concentration exceeds the limit The alarm can be stopped by pressing any button except SET It will also stop when readings fall below the set value P3 0 Set Temperature Units Fr...

Страница 6: ... cali bration turn off the meter at any time NOTE Ensure the batteries are at full power during the cali bration to prevent interruption or a failed calibration RH Calibration Humidity calculations on the meter can be recalibrated using 33 and 75 salt solutions Ambient conditions should be at 25 C or 77 F with stable humidity Plug the sensor probe into a 33 salt bottle Hold down the CAL ESC and DP...

Страница 7: ...t the left top Slow sensor response Check whether the air flow channels on the back of the meter are blocked Error messages E01 CO2 sensor damaged E02 The value is under range E03 The value is over range E04 Error code for dew point or wet bulb accompanying a primary sensor failure code E07 Too low voltage to measure CO2 Replace batteries E11 Retry humidity calibration E17 Retry CO2 calibration E3...

Страница 8: ...d to be free from defects in material or work manship for one year from the date of purchase During the warranty period Spectrum will at its option either repair or replace products that prove to be defective This warranty does not cover damage due to improper installation or use lightning negligence accident or unau thorized modifications or to incidental or consequential damages be yond the Spec...

Отзывы: