Power ON/OFF  

The meter automatically powers on when the trigger is pulled. Pull  
the trigger to take a reading. Read the measured temperature on  
the LCD. The meter powers OFF automatically approximately 6  
seconds after the trigger is released.  
Selecting Temperature units (C/F)  
Select temperature units by pulling and holding the trigger, then  
press the C/F key until the desired units are shown on the LCD.  

Data Hold  

The meter automatically holds the last temperature reading on the  
LCD for 6 seconds after the trigger is released.   

Backlight LCD  

Select backlight by first pulling the trigger and then pressing the   
BACKLIGHT key. Repeat the procedure to turn the backlight off.  

Laser Pointer  

To turn the laser pointer ON, press the LASER key while pulling the  
trigger. Repeat the procedure to turn the laser OFF.  

Measurement Considerations  

Point the IR Sensor toward the object whose temperature is to be  
measured.  The meter automatically compensates for   
deviations from ambient temperature.  However, the meter will take  
up to 30 minutes to adjust to wide ambient temp. changes.  This  
results from the need to acclimate the IR sensor.  
  

Battery Replacement  

A flashing display indicates that the battery voltage has fallen into  
the critical region (6.5 to 7.5 V). Reliable readings can be obtained  
for several hours after the first appearance of the low battery  
indication. Open the Battery Compartment and remove the battery,  
then install a new 9V battery and reinstall the cover.  
  

Maintenance  

Case cleaning  
Use caution with a damp cloth to clean the exterior housing, ensure  
no water or soap is allowed inside the meter and on the infrared  
lens.  
Lens cleaning  
Keep the lens clean to ensure reading accuracy. Use low pressure  
compressed air to remove any particles on the lens. If the  
contamination can not be removed with air, use a soft cotton swab.  
Any swab should be slightly damp, and very light pressure should  
be applied to the lens.   
Do not use solvents to clean the lens.  