Setting meter for Auto Temperature Compensation (sect. 3.4.4 of manual)
1. From Measurement mode, press the SET button until you’re in
Temperature Compensation Setting Mode (fig. 9).
2. Press ENTER until the screen displays “Atc”.
3. Press MEAS to return to measurement mode.
Figure 9
Set Cell Constant (sect. 2.8 of manual)
1. From Measurement mode, press the MODE button until you’re in COND measurement mode
(fig. 10).
2. Press CAL to enter CELL SET mode (fig. 11).
3. Press ENTER until the screen displays CELL x 1(fig. 12).
3. Press the arrow buttons to adjust the cell constant to the value written on top of the probe.
Figure 12
Figure 11
Figure 10
Calibrating EC (sect. 2.8 of manual)
1. Immerse electrode in EC calibration solution. Probe should be immersed above the hole.
2. From COND Measurement mode (this can be selected by pressing
the MODE key), press CAL to enter CELL SET mode.
3. Press MODE to enter calibration mode (fig. 13).
4. Press the arrow buttons to adjust the display to the expected value.
5. Press CAL to initiate calibration.
6. The HOLD icon will stop blinking when calibration is complete.
7. Press MEAS to return to measurement mode.
Figure 13