Setting the pH7 standard solution mode (sect. 3.4.3 of manual) 
The meter allows for calibration at a pH of either 6.8 (NIST), 7.0 (USA) or a customer-
specified value.  Spectrum ships the meter with pH 7.0 solution.  To confirm the meter is in 
USA mode: 
1. With the meter in measurement mode, press the SET key to enter the  pH Buffer selection 
mode. 
2. Pressing the ENTER key will toggle you between the 3 modes.  Select USA mode. 

 
3. Press the MEAS 
key to return to 
measurement 
mode. 

 

D-54 pH/EC Meter (item 2172) 

Quick Start Guide

 

 

pH 

 
Probe preparation  (sect. 6.1. of manual) 
The pH probe should  be connected to channel 1 of the D-54 Reader (see fig. 1).  The probe 
should always have a sufficient quantity of internal filling solution (item # N360647

 

).  This 

solution should be changed every three to six months to maintain probe performance.  The old 
solution should be removed with a syringe and new solution added until it is completely filled, 
leaving few or no bubbles (fig. 2).   