The meter and probe are factory calibrated. How-
ever, to ensure accuracy, the meter should be cali-
brated at least once a month. The meter is cali-
brated to an EC of 2764
µ
S/cm (2.76 mS/cm) cali-
bration solution. Periodically, the calibration of the
meter can be checked by immersing the sensor in
this solution. This solution should not be re-used at
a later date.
Calibration Procedure
1.) Open battery compartment lid on the top of the
meter to access the small white Increment (INC) and
Decrement (DEC) calibration keys. When looking at
the battery compartment with the calibration keys
above the batteries, the DEC key will be to the left.
Probe Calibration
DEC
key
INC
key