Return Authorization Procedure
A defective PRO-Alert™ Refrigerant Leak Detector, or a defective replacement part or
accessory, should be returned to your wholesaler for warranty evaluation. If you have any
questions, please contact the Customer Service Department at Spectronics Corporation.
NOTE
: Do not return your defective unit directly to the factory without first contacting
Spectronics for a Return Authorization Number.
Replacement Parts and Accessories
Replacement parts and accessories for your PRO-Alert™ Refrigerant Leak Detector are
available through the same dealer from which you bought the instrument.
Replacement sensor . . . . . .TP03
Tip filters, package of 20. . . TP04
Warranty and Liability
Your PRO-Alert™ Refrigerant Leak Detector is of the finest design and craftsmanship. As
such, Spectronics Corporation warrants that when delivered it will be free from defects in
manufacture.
Spectronics Corporation's obligation under this warranty is limited to repairing or replacing,
at the option of Spectronics Corporation, any part(s) of the product which, if properly
installed, used and maintained, proves upon factory examination to have been defective in
materials or workmanship within two (2) years from the date of delivery.
Spectronics Corporation does not warrant items that deteriorate under normal use,
including power cells, sensors and filters. In addition, Spectronics Corporation does not
warrant any instrument that has been subjected to misuse, negligence or accident, or has
been repaired or altered by anyone other than Spectronics Corporation.
If you do not hear the
motor, please contact
the Customer Service
Department at
Spectronics Corporation.
2c.The sensor cover is not
sealing.
2c. Make sure the sensor
cover is properly installed.
See step 5 on page 4.
3. Will not power up.
3a. Batteries are worn out.
3a. Install a new set of
batteries. See page 3-4.
3b. Batteries have been
improperly installed.
3b. Check battery
installation as shown
in Figure 1, on page 4.
4. False alarms — the
PRO-Alert™ alarms
when
.
the probe is moved
or bumped.
4a. Sensor leads are bent.
4a. Remove the sensor and
inspect the leads.
Straighten the leads with
needle nose pliers, if
necessary, and reinstall
the sensor.
4b. Moisture was absorbed
by the sensor during a
long period without use.
4b. Run the PRO-Alert™ for
at least 20 minutes. The
absorption of moisture
does not affect the life
or sensitivity of the sensor.
Содержание Pro Alert TP-9360
Страница 1: ...1 OPERATING MANUAL TP 9360 4 22 AM06026ML 7 Printed in U S A ENG 1 FR 8 DE 24 ES 16 ...
Страница 8: ...8 TP 9360 Français ENG 1 FR 8 DE 24 ES 16 MANUEL D UTILISATION ...
Страница 16: ...16 TP 9360 Español ENG 1 FR 8 DE 24 ES 16 MANUAL DE INSTRUCCIONES ...
Страница 24: ...24 TP 9360 Deutsche ENG 1 FR 8 DE 24 ES 16 BENUTZERHANDBUCH ...
Страница 32: ...32 ...