4.3 Manually screw the regulator onto the cylinder valve (1). Tighten connection 

gas-tight by using an appropriate spanner. Do not use any extension to the 
spanner to increase momentum! In case of flammable and toxic gases attach a 
relief line to the relief valve (4) and ensure a safety gas discharge. For cylinder 
regulators with external purge gas connect the purge gas line.

4.4 Carry out the connections between the cylinder pressure regulator and the point 

of use. Never allow cylinder pressure regulator to regulate into the open 
atmosphere.
Wind PTFE tape (Art. No. 0321422) clockwise around NPT-screwed 
connections (5 to 10 turns). Check all connections for leaks.

4.5 Release the pressure regulator spring with the hand knob (6) - open the cylinder 

valve (1) (High pressure gauge 3 indicates the cylinder pressure) - Adjust the 
required outlet pressure with hand knob (6). Correct the pressure setting in case 
of a decrease of pressure.

!

1

Cylinder valve to DIN 477 or the relevant national standard

2. Regulator connection to

DIN 477 or the relevant national standard

3. High-pressure gauge 
4. Relief valve
5. Low-pressure gauge
6. Hand knob

 4. Start-up

4.1 Before starting read the specifications of this instruction for use and observe it 

while working.

4.2 Check, that the cylinder valve thread, pressure regulator, connector and 

connection seals are without any damage (blow through if necessary).

In case of a damage, the cylinder pressure regulator must not be connected.

 3. Labelling

1  Type
2  Gas type
3  Article no.
4  Max. inlet press.
5  Max. outlet press.
6  Date of manufacture

0

0,2

1

2

3

10

0,3

0,5

4

5

0

0,4

0,8

1,2

1,6

3

Flow rate Q [m /h air]

Outlet pressure P

bar]

H[

Inlet pressure P  = 4 bar

V

5. Operation and maintenance

5.1 Before commisioning check on the nameplate if the regulator is suitable for the 

gas type. The designated gas type has to be marked on the nameplate using gas 
type stickers or by writing the gas type into the space provided on the 
nameplate. 

FE 121 - O

P1: 25 bar

P2: 1,5 bar

717. …
0601

1

2

3

4

5

6

1

2

3

4

5

6

SW 27

3

2. Safety instructions

- Guidelines for laboratories (Zh 1/119)
- Technical regulation (liquified pet. gas)

In case of flammable gases, thedischarge of the gas from the relief valve (4) 
must be ensured. For this use connector 1/8-NPT/6mm pipe (art. no. 0291677).

No adapters must be used between the gas cylinder valve and the pressure 
regulator.

2.1

All items of information marked with   are valid as special safety instructions.

2.2

These pressure regulators adhere to state-of-the-art technology and 

    

to the demands of the exsiting standards and regulations.

2.3

Changes  or  modifications  are  not  allowed  to  be  made  to  the  pressure 
regulator without the prior consent of the manufacturer.

2.4

The result of improper handling and improper use as intended can involve 
risks for the user and other persons as well as damage to the device.

2.5

The equipment must be operated by suitable trained personnel only.

2.6

Regulations to be adhered to:
- BGV A1 (VBG 1), “General specifications”
- BGV B6 (VBG 61), “Gases"
- BGV B7 (VBG 62), “Oxygen”

Special attention has to be paid to the country specific laws, regulations 
and procedures concerning the use of this type of equipment. 

2.7

Use only for gas types the pressure regulator is labelled for (see item 3).

2.8

2.9

Do not use at temperatures below -30°C or above +60°C.

2.10 The valve has always to be opened slowly!   
2.11 All  parts  coming  into  contact  with  oxygen  must  be  kept  in  oil-free  and 

grease-free condition.

Fire or explosion hazard!

2.12 Smoking or open fire (e.g. candles) in the vicinity of the gas supply 

system is strictly prohibited.

Fire and explosion hazard!

2.13 Protect gas cylinders against falling.
2.14

!

!

!

!

!

!