![Spectron FS4 Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/spectron/fs4/fs4_user-manual_1344164019.webp)
Spectron Gas Control Systems GmbH
Installation | 5
GES_FS4_0520
19
WARNING
Cross-contamination
Cross-contamination of the control air with process gases can result in the control unit being
damaged.
a) Ensure that cross-contamination of the control air line is prevented.
The solenoid valves require a supply of clean dry air or nitrogen filtered to 10μm, at max. 5.5
bar. This is connected via a 4mm push-in fitting on the base of the device as indicated on the la-
bel. Any exhaust from the solenoid valves is vented internally and excess pressure build-up is
vented to atmosphere via an IP65 breather on the base of the unit.
Содержание FS4