Spare parts and consumables
— 28
SPECTRO iSORT — 23.08.2013
Fault
Possible cause
Remedy
The result falls below the
stored reference element
limit values (error
message).
The adapter in the "Arc"
mode is reversed.
Turn the adapter so that one
opening faces the optical fiber
window.
The adapter is not locked
in properly up to the stop.
Check adapter seat and correct it
if necessary.
The electrode gap is not
correct.
Adjust the electrode gap using
the spacer.