Operation
— 32
SPECTROMAXx — 21.02.2020
6.4.2
Start the measurement
Information
Use the electrode brush in order clean the electrode tip after each measurement.
To measure the probe, proceed as follows:
1. Start the required measuring program using function key F10.
2. Put the sample in position.
The sample must fully cover the opening in the spark stand plate.
The sample clamp must have contact with the sample (the safety device
is activated).
3. Use "Start" or the respective function key to start the measurement.
The measurement is fully automated.
The measurement is finished automatically.
The measurement results are displayed on the screen.
4. Perform further measurements as needed.
Use the electrode brush in order clean the electrode tip after each
measurement.
Rotate the sample slightly for each further measurement. The respective
burn spots must not overlap.
5. Use "Start" or press the respective function key to start the series of
measurement.
The measurement process is finished.